l Operation Panel I/F
• LED ON/OFF control ................................................................×11
• Switch .......................................................................................×3
• Buzzer ......................................................................................×1
l Others
For Analysis
• Fuse disconnection detection circuit
• Board version information is controlled by the difference of pullup/down
resistance inplementation
016-201-06E
01.27.2012 FM5920
3.3.3
RLY Board
n Outline
The RLY board performs switching between the head power supply and the resistance
measurement.
+14.8V
Power
supply
Resistance
measurement
SND32A
Head power supply/resistance measurement switching signal
DRYPIX Lite Service Manual
Head
RLY32A
MD-18
D0011.ai
MD-18