3.
PREVENTIVE MAINTENANCE
3.1
Checking the Error Logs
1Y
2Y
3Y
4Y
5Y
Check whether errors have occurred using PC-Utility. If errors have occurred, correct
them appropriately. If necessary, save the error logs and delete those error logs.
Procedure
l Checking Error Logs
1.
Turn ON the power of the equipment.
2.
Start PC-Utility during initialization.
3.
Check the error log.
{IN (DRYPIX
Lite):10.2.1_Checking the Error Logs}
l Saving Error Logs
Save the log files in the PC for servicing as necessary.
l Deleting Error Logs
1.
Clear the log file.
{IN (DRYPIX
Lite):10.2.2_Deleting Error Logs}
016-201-06E
01.27.2012 FM5920
3.2
Printing/Checking Conveyance/Checking
Images
1Y
2Y
3Y
4Y
5Y
Print 17-steps pattern and Grid & Flat pattern using PC-Utility, and check the density
correction function, recorded image format, and there are no unevenness on the
images and scratches.
Procedure
1.
The data of the 17-steps pattern measured by the density
measurement section is displayed on the PC-Utility screen. Check
from this data that the density value is within the specification.
{IN (DRYPIX
Lite):8.2_Checking Density}
2.
Print Grid & Flat pattern, and check the operations of the mechanism
system and conveyor system, recorded image format, and also check
for unevenness and scratches on the images.
{IN (DRYPIX
Lite):8.3_Checking Recording Image Format, and for Image
Unevenness and Scratches}
DRYPIX Lite Service Manual
PM-5
PM-5