Equipment Installation Space; Drypix 2000; Space Required For Maintenance Work - FujiFilm DRYPIX 2000 Service Manual

Medical dry imager
Table of Contents

Advertisement

n Grounding Resistance
100 Ω or less
n Overload Protection
10 A
n Power Cable
CAUTION
If the power cable is not packed together with the equipment, be sure to use a
power cable that complies with the requirements stated below.
Using the cables not corresponding to the conditions below may cause an
electric shock or a fire.
l For U.S. and Canada
• UL-listed detachable power supply cable
• Hospital-grade Type Plug
• Cable wire diameter:
16 AWG or larger, 3-conductor
• Rated voltage:
125 VAC or higher
• Cable type:
SJT
l For Europe
• Cable certified by a country in which the equipment is to be installed
• Cable wire diameter:
1.0 mm
• Rated voltage:
250 VAC or higher
• Cable type:
H05VV-F
n Network Cable
{IN (DRYPIX
2000):6.1_Connecting the Network Cable}
{IN (DRYPIX
Lite):6.1_Connecting the Network Cable}
016-201-07E
04.27.2012 FM6040
2
or larger, 3-conductor
1.7

Equipment Installation Space

<NOTE>
The following space does not include space for people.
1.7.1

DRYPIX 2000

n Installation Space
Front: Magazine S; More than 400 mm
Magazine L; More than 500 mm
Rear: More than 150 mm
Top:
More than 500 mm
Left:
More than 50 mm
Right: More than 50 mm
n Space Required for Maintenance Work
Front: Magazine S; More than 400 mm
Magazine L; More than 500 mm
Rear: More than 200 mm
Top:
More than 700 mm
Left:
More than 100 mm
Right: More than 100 mm
DRYPIX 2000 / DRYPIX Lite Service Manual
-10
SPECIFICATIONS
-10
SPECIFICATIONS

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Drypix lite

Table of Contents