[4-5-3] Clr Colle. T.>
<FUNCTION/PURPOSE>
Command for linearly initializing densitometer calibration tables.
<OPERATIONS>
1.
[Are you sure?] →
→ T he calibration tables are cleared.
When it completes normally, "Completed." is displayed.
<IMPLEMENTATION OF SETTINGS>
The settings made will be effective after they are executed.

[4-5-4] DM Sen. Moni.

<FUNCTION/PURPOSE>
Command for monitoring whether the density sensor of the internal densitometer is
operating normally.
<OPERATIONS>
1.
The state of the density measurement section sensor is displayed in a
constant cycle.
Shield the density measurement section sensor with the film, and
check that low values are displayed for high density areas of the film
and high values for low density areas.
016-201-01E
06.30.2006 FM4924
[4-6]
Densito.Lamp>
<FUNCTION/PURPOSE>
Command for turning ON/OFF the density measurement LED of the built-in
densitometer.
<OPERATIONS>
1.
Select [ON].
→ T he density measurement LED turns ON.
2.
Select [OFF].
→ T he density measurement LED turns OFF.
DRYPIX 2000 Service Manual
MU-32