[PC8] Head Check
[PC8-1] Temperature
<FUNCTION>
Command for performing temperature monitoring of each thermistor.
<REMARKS>
The displayed temperature is ten times the value and can be displayed to the first
decimal point.
<NOTE>
After completing monitoring, execute [OFF] to end communication with the equipment.
If communication is not ended, it may not be possible to execute other menus.
[PC8-2] Fan Operation
<FUNCTION>
Command for tuning ON/OFF the head cooling fans (FANP1, FANP2).
<OPERATIONS>
1.
Select operation.
• [OFF]
: Fan stops
• [ON]
: Fan rotates
• [Usually]
: Fan operates normally
2.
[Execute]
→ T he fan rotates.
3.
[Usually] → [Execute]
→ T he fan returns to normal operations.
016-201-06E
01.27.2012 FM5920
[PC8-3] Head Voltage.
<FUNCTION>
Command for turning ON/OFF the head power (+14.8V) when adjusting the head
power voltage.
<OPERATIONS>
1.
[ON] → [Execute]
→ 1 4.8V is output.
2.
Adjust the head power voltage.
3.
[OFF] → [Execute]
→ O utput of 14.8V stops.