3.

CONNECTABLE PERIPHERALS

Accessory equipment connected to the analog and digital interfaces must be certified
according to the respective IEC standards (i.e. IEC 60950 for data processing
equipment and IEC 60601-1 for medical equipment). Furthermore all configurations
shall comply with the system standard IEC 60601-1-1. Everybody who connects
additional equipment to the signal input part or signal output part configures a medical
system, and is therefore responsible that the system complies with the requirements
of IEC 60601-1-1. If in doubt, consult the technical services department or your local
representative.
016-201-01E
03.24.2006 FM4903
4.

CLASSIFICATION

1)
According to the type of protection against electrical shock
CLASS 1 EQUIPMENT
2)
According to the degree of protection against electrical shock
NO APPLIED PART
3)
Protection against harmful ingress of water
IPXO
4)
According to the degree of safety of application in the presence of
a flammable anesthetics mixture with air or with oxygen or nitrous
oxide.
Equipment not suitable for use in the presence of a flammable anesthetics
mixture with air or with oxygen or nitrous oxide.
5)
According to the mode of operation
CONTINUOUS OPERATION
DRYPIX 2000 Service Manual
-7
SAFETY PRECAUTION
-7
SAFETY PRECAUTION