n Reinstalling Procedure
<INSTRUCTIONS>
When performing the following procedure, move the suction cup arm to the home
position and then begin work.
If the suction cup arm assembly is not at the home position during the work, a phase
shift occurs.
<REMARKS>
The removing procedure of the removal drive cam is common between the left and
right sides.
This section describes the removing procedure for the left side only.
1.
Reinstall the crank and the removal drive cam as well to the removal
section.
Crank
Removal drive cam
016-201-06E
01.27.2012 FM5920
Bearing
F0074.ai
2.
Trial fit the Gear 1 and the Gear 2 to adjust the position of the Gear 3.
#1: Align the hole on the removal drive can with the hole on the Gear 1.
#2: Align the rib on the Gear 1 with the l mark on the Gear 2.
#3: Align the rib on the Gear 2 with the rib on the Gear 3.
Removal drive cam
Removal drive cam
Removal drive cam
Removal drive cam
#1
#1
Gear 1
Gear 1
3.
Remove the trial fit Gear 1 and the Gear 2.
DRYPIX Lite Service Manual
MC-18
Gear 3
Gear 3
Gear 2
Gear 2
#2
#2
#3
#3
MC-18
F0076.ai