FujiFilm DRYPIX 2000 Service Manual page 462

Medical dry imager
Table of Contents

Advertisement

[1-1-6-2] Init.Trip-cnt>
<FUNCTION/PURPOSE>
Command for resetting the user counter for each magazine.
<OPERATIONS>
1.
Select the number of the magazine to be reset.
→ The confirmation message "Are you sure?" appears.
2.
Press the <ENTER> key.
→ The user counter is reset.
<IMPLEMENTATION OF CHANGES>
The changes made will be effective immediately.
[1-1-6-3] Set Odo-cnt>
<FUNCTION/PURPOSE>
Command for changing the counter value of the total film counter.
<OPERATIONS>
1.
Select the number of the magazine to change the counter value.
2.
Enter the counter value.
<REMARKS>
Input range: 0 to 999999 (6 digits)
<IMPLEMENTATION OF CHANGES>
The changes made will be effective immediately.
016-201-04E
02.15.2007 FM5075
[1-1-6-4] Set Rem.Films>
<FUNCTION/PURPOSE>
Command for changing the number of remaining films on the magazines loaded in the
equipment.
<OPERATIONS>
1.
Select the magazine slot.
2.
Enter the number of remaining films.
<REMARKS>
Input range: 0 to 101 (fixed at three digits) ("101" should be entered only
immediately after films have been loaded.)
<IMPLEMENTATION OF CHANGES>
The changes made will be effective immediately.
DRYPIX 2000 Service Manual
MU-17
MU-17