Configure alert handling

The alert handling parameters control the transmitter's response to process and device
conditions.
Configuration and Use Manual
Option Description
Enabled Default.Operator can access the maintenance section of the off-line menu. This
access is required for configuration and calibration, including Known Density Veri-
fication.
Disabled Operator cannot access the maintenance section of the off-line menu.
To control operator access to the alert menu, enable or disable Alert Menu.
Option Description
Enabled Default. Operator can access the alert menu. This access is required to view and
acknowledge alerts, but is not required for Known Density Verification, configura-
tion, or calibration.
Disabled Operator cannot access the alert menu.
Note
The transmitter status LED changes color to indicate that there are active alerts, but does not
show specific alerts.
To require a passcode for access to the off-line menu, enable or disable Off-Line
Password.
Option
Description
Enabled
Operator is prompted for the off-line passcode at entry to the off-line menu.
Disabled
Default. No passcode is required for entry to the off-line menu.
Set Off-Line Password to the desired value.
The default value is 1234. The range is 0000 to 9999.
Tip
Record your passcode for future reference.