The status LEDs use colors and flashing to indicate device status.
Table 14-1:
Transmitter status reported by status LED
LED state
Description
Green
No alerts are active.
Yellow
One or more low-severity alerts are active.
Flashing yellow
Calibration in progress, or Known Density Veri-
fication in progress.
Red
One or more high-severity alerts are active.
Related information
14.6

Status alerts, causes, and recommendations

Table 14-2:
Status alerts, causes, and recommendations
Alert number
Alert title
A001
EEPROM Error
A002
RAM Error
152
A status LED on the board, beneath the meter housing cover
View and acknowledge status alerts
Possible causes
The transmitter has detec-
ted a problem communicat-
ing with the sensor.
The transmitter has detec-
ted a problem communicat-
ing with the sensor.
Recommendation
Continue with configuration or process meas-
urement.
A low-severity alert condition does not affect
measurement accuracy or output behavior.
You can continue with configuration or proc-
ess measurement. If you choose, you can iden-
tify and resolve the alert condition.
The low-severity alert condition does not af-
fect measurement accuracy or output behav-
ior. You can continue with configuration or
process measurement. If you chose, you can
identify and resolve the alert condition.
A high-severity alert condition affects meas-
urement accuracy and output behavior. Re-
solve the alert condition before continuing.
Recommended actions
Cycle power to the meter.
Contact Micro Motion.
Cycle power to the meter.
Contact Micro Motion.
®