Populate the equations:
Solve for A:
Solve for B:
Concentration slope (A): 1.00029
Concentration offset (B): 0.00551
Related information
13.7

Set up user-defined calculations

ProLink III
Field Communicator Configure > Manual Setup > Measurements > Optional Setup > User-Defined Calcula-
Overview
User-defined calculations are used to enhance measurement or to adapt the meter to
special process conditions.
A user-defined calculation allows you to create a new process variable by inserting
constants and existing process variables into an equation. The output of the equation is
the new process variable. Depending on your meter, either two or three equations are
available.
Procedure
1.
2.
140
50.00 = 1.00029 × 49.98 + �
Adjust concentration measurement with Trim Offset
Device Tools > Configuration > User-Defined Calculations
tions
Select the user-defined calculation that you want to use.
If you selected User-Defined Calculation 1:
50 = � × 49.98 + �
16 = � × 15.99 + �
50.00 − 16.00 = 34.00
49.98 − 15.99 = 39.99
34 = � × 33.99
� = 1.00029
50.00 = 49.99449 + �
� = 0.00551
®