Configure a concentration measurement matrix using
the Field Communicator
A concentration measurement matrix defines the relationship between density and
concentration for your process gas. You can configure up to six matrices.
Previous:
Prerequisites
You must know the primary and secondary constituents of your process gas, and the
density of each constituent in pure form.
Procedure
1.
2.
3.
4.
5.
6.
7.
8.
Next:
8.4.3
Select the active concentration matrix using the
Field Communicator
You must select the concentration matrix to be used for measurement. Although the
transmitter can store up to six concentration matrices, only one matrix can be used for
measurement at any one time.
Previous:
1.
2.
Configuration and Use Manual

Configure a concentration measurement matrix using the Field Communicator

Enable the concentration measurement application using the Field Communicator
Choose Configure > Manual Setup > Measurements > Conc Measure (CM) >
Configure Matrix.
Set Matrix Being Configured to the matrix you want to configure.
Enter a name for the matrix.
Set Concentration Units to the label to used for the measurement unit.
This selection does not affect measurement. It only selects a label.
If you set Concentration Units to Special, enter a string for the custom label.
Choose Enter Matrix Data.
Enter the density of the primary constituent of your process gas, in its pure form.
Enter the density of the secondary constituent of your process gas, in its pure form.

Select the active concentration matrix using the Field Communicator

Configure a concentration measurement matrix using the Field Communicator
Choose Configure > Manual Setup > Measurements > Optional Setup >
Concentration Measurement > CM Configuration.
Set Active Matrix to the matrix you want to use.
Configure process measurement using the Field Communicator
83