Configure process measurement using the Field Communicator
If mA Output Process Variable is set to Specific Gravity, Molecular Weight, or Relative Density, and
both Damping and Added Damping are set to non-zero values, density damping is applied
first, and the added damping calculation is applied to the result of the first calculation. This
value is reported over the mA output.
8.2
Configure temperature measurement using
the Field Communicator
The temperature measurement parameters control how temperature is measured and
reported.
8.2.1
Configure Temperature Measurement Unit using the
Field Communicator
Field Communicator Configure > Manual Setup > Measurements > Temperature > Temperature Unit
Overview
Temperature Measurement Unit specifies the unit that will be used for temperature
measurement.
Procedure
Set Temperature Measurement Unit to the option you want to use.
The default setting is Degrees Celsius.
Important
If you plan to use an external temperature device, you must set Temperature Measurement Unit to
the unit used by the external device.
Options for Temperature Measurement Unit
The transmitter provides a standard set of units for Temperature Measurement Unit.
Different communications tools may use different labels for the units.
74

Configure Temperature Measurement Unit using the Field Communicator

(Section 8.2.1)
Configure Temperature Damping using the Field Communicator
®
Micro Motion
Gas Specific Gravity Meters (SGM)
(Section 8.2.2)