12
Transmitter operation
Topics covered in this chapter:
12.1

Record the process variables

Micro Motion suggests that you make a record of specific process variable measurements,
including the acceptable range of measurements, under normal operating conditions. This
data will help you recognize when the process or diagnostic variables are unusually high or
low, and may help you diagnose and troubleshoot application issues.
Procedure
Record the following process and diagnostic variables, under normal operating conditions.
Variable
Specific Gravity, Molecular
Weight, or Relative Density
Line Temperature
Line Pressure
Sensor Time Period
Pickoff Voltage
Drive Gain
12.2

View process variables and diagnostic variables

Process variables provide information about the state of the process fluid. Diagnostic
variables provide data about meter operation. This information can be used to understand
and troubleshoot your process.
12.2.1

View process variables and other data using ProLink III

Monitor process variables, diagnostic variables, and other data to maintain process quality.
Configuration and Use Manual
Record the process variables
View process variables and diagnostic variables
View and acknowledge status alerts
Typical average
Measurement
Typical high
Transmitter operation
Typical low
123