Enable Access To The Off-Line Menu Of The Display; Disable Hart Security - Emerson Micro Motion SGM Series Configuration And Use Manual

Gas specific gravity meters
Table of Contents

Advertisement

Table 4-2:
Default variable
Primary Variable (PV), mA1
Secondary Variable (SV),
mA2
Tertiary Variable (TV)
Quaternary Variable (QV)
4.2
Enable access to the off-line menu of the
display
ProLink III
Field Communicator Configure > Manual Setup > Display > Display Menus > Offline Menu
Overview
By default, access to the off-line menu of the display is enabled. If it is disabled, you must
enable it if you want to use the display to configure the transmitter.
Restriction
You cannot use the display to enable access to the off-line menu. You must make a connection from
another tool.
4.3

Disable HART security

If you plan to use HART protocol to configure the device, HART security must be disabled.
HART security is disabled by default, so you may not need to do this.
Prerequisites
Configuration and Use Manual
SGM default variables
Output option A
Sample Temperature
Time Period B
Drive Gain
Device Tools > Configuration > Transmitter Display > Display Security
Strap wrench
3 mm hex key
Introduction to configuration and commissioning
Output options B and C
Sample Temperature
Specific Gravity for
Sensor Time Period
Calibration Set 1
Molecular Weight for
Calibration Set 1
Relative Density
Drive Gain
Specific Gravity for Calibration Set 1
Molecular Weight for Calibration Set
1
Relative Density
23