Troubleshooting
Table 14-2:
Status alerts, causes, and recommendations
Alert number
Alert title
A014
Transmitter
Failure
A016
Sensor Temper-
ature (RTD)
Failure
A020
Calibration Fac-
tors Missing
A021
Transmitter/
Sensor/Soft-
ware Mismatch
A029
Internal Elec-
tronics Failure
A030
Incorrect Board
Type
A033
Insufficient
Pickoff Signal
A037
Sensor Check
Failed
A038
Time Period
Signal Out of
Range
154
Possible causes
Many possible causes.
The value computed for the
resistance of the line RTD is
outside limits.
Some calibration factors
have not been entered or
are incorrect.
The configured board type
does not match the physical
board.
This can indicate a loss of
communication between
the transmitter and the dis-
play module.
The loaded software is not
compatible with the pro-
grammed board type.
The signal from the sensor
pickoff(s) is insufficient. This
suggests that the sensor
tubes or vibrating elements
are not vibrating. This alert
often occurs in conjunction
with Alert 102.
Known Density Verification
failed.
The time period signal is
outside the limits for the
sensor type.
(continued)
Recommended actions
Ensure that all wiring compartment covers are in-
stalled correctly.
Ensure that all transmitter wiring meets specifica-
tions and that all cable shields are properly termi-
nated.
Check the grounding of all components. See
Section
14.3.
Evaluate the environment for sources of high elec-
tromagnetic interference (EMI) and relocate the
transmitter or wiring as necessary.
Contact Micro Motion.
Check your process against the values reported by
the device.
Contact Micro Motion.
Verify all of the characterization or calibration pa-
rameters. See the sensor tag or the calibration
sheet for your meter.
Perform site calibration procedures.
Verify all of the characterization or calibration pa-
rameters. See the sensor tag or the calibration
sheet for your meter.
Ensure that the correct board is installed.
Cycle power to the meter.
Replace the display module.
Contact Micro Motion.
Contact Micro Motion.
Check for foreign material in the process gas or
fluid, coating, or other process problems.
Check for fluid separation by monitoring the den-
sity value and comparing the results against ex-
pected density values.
Ensure that the sensor orientation is appropriate
for your application. Settling from a two-phase or
three-phase fluid can cause this alert.
Check the subtest results and perform the recom-
mended actions.
Retry the test.
Contact Micro Motion.
Check your process conditions against the values
reported by the device.
®