Important
Whenever you change a discrete output parameter, verify all other discrete output parameters
before returning the meter to service. In some situations, the transmitter automatically loads a set of
stored values, and these values may not be appropriate for your application.
10.3.1

Configure Discrete Output Source

ProLink III
Field Communicator Configure > Manual Setup > Inputs/Outputs > Discrete Output > DO Source
Overview
Discrete Output Source controls which device condition or process condition is reported
via the discrete output.
Procedure
Set Discrete Output Source to the desired option.
The default setting for Discrete Output Source is Fault.
Options for Discrete Output Source
Table 10-4:
Options for Discrete Output Source
Option
ProLink III
Enhanced Event 1–5
Enhanced Event 1
Enhanced Event 2
Enhanced Event 3
Enhanced Event 4
Enhanced Event 5
Calibration in Progress
Calibration in Progress
Fault Indicator
Fault (default)
Important
This table assumes that Discrete Output Polarity is set to Active High. If Discrete Output Polarity is set
to Active Low, reverse the voltage values.
Configuration and Use Manual
Device Tools > Configuration > I/O > Outputs > Discrete Output
Label
Field Communicator
Enhanced Event 1
Enhanced Event 2
Enhanced Event 3
Enhanced Event 4
Enhanced Event 5
Calibration in Progress
Fault
Integrate the meter with the control system
Discrete output volt-
State
age
ON
Site-specific
OFF
0 V
ON
Site-specific
OFF
0 V
ON
Site-specific
OFF
0 V
103