Invert Image; Auto Contrast; Adjust Image - Thermo Scientific myECL Imager Manual

Table of Contents

Advertisement

Invert Image

The Invert button is a useful tool for viewing faint gel bands. This tool inverts the
standard black background with white highlights to a white background with black
highlights.
Inverted UV Image
Original UV Image
1. In the active live image, select the Invert button from under the time keypad.
2. The inverted image can be switched back to the standard image by again
selecting the Invert button.

Auto Contrast

Selecting the Auto Contrast button increases the contrast between the signal and
the background to increase visibility. This adjustment can be used only for viewing or
printing of the image. The contrast adjustment is NOT saved with the image file.
Note Auto Contrast is disengaged when the Adjust Image button is selected.

Adjust Image

Select the Adjust Image button to modify the Black, White and Gamma levels of the
image to increase sample visibility. This adjustment is only used for viewing or
printing the image. Image adjustments are NOT saved with the image file.
Select the Reset button to restore Black, White and Gamma to original levels. Select
the Close button to return to the previous screen.
Note Adjust Image is disengaged when the Auto Contrast button is selected.
50
ECL Imager User Manual
Thermo Fisher Scientific
MY

Advertisement

Table of Contents
loading

Table of Contents