Multiple Image Acquisition

36
ECL Imager User Manual
MY
1. Select the Chemi tab.
2. (Optional) Select the Live Camera button if the live image is not displayed.
3. The epi-white lights and live image display are activated. Open the imager door,
pull out the transilluminator and install the chemiluminescence exposure screen
(black) on the transilluminator glass.
4. Insert and center the sample on top of the chemiluminescence exposure screen
(black) using the live image display for reference.
5. Once the sample is centered, close the imager door.
6. While viewing the live image, adjust the settings for Focus, Digital Zoom and
Resolution/Sensitivity in the Camera Settings menu, if needed.
Note Camera Settings are available only with the live image display.
7. Select a Pre-set or User-defined Multiple Exposure Time.
Note A Stop button appears over the keypad once the Acquire button is
selected. Image acquisition can be stopped by selecting the Stop button.