Image Acquisition –
Reflected Light

Pre-set Exposure Time

53
ECL Imager User Manual
MY
1. Select the Vis tab on the touchscreen. Changing to Vis mode automatically turns
on the UV lights within the transilluminator and the epi-white lights.
2. (Optional) Select the Live Camera button and enable the UV button if the live
image is not displayed.
3. Open the imager door, confirm that the UV light is off, and install the white light
conversion screen on top of the transilluminator glass.
4. Insert and center the sample on top of the white light conversion screen using
the live image display for reference.
5. Once the sample is centered, close the imager door. The UV light turns on.
6. Select the UV light button on the right side of the screen to switch OFF the UV
light. This produces only reflected light.
7. While viewing the live image display, adjust the Camera Settings for Focus,
Digital Zoom, Resolution/Sensitivity, Filter and Aperture to produce the desired
image, if needed.
8. Select a Pre-set or Custom exposure time.
1. Select one of the Pre-set exposure time buttons using the keypad.
2. Select the Acquire button to capture an image.