Pedestal Image Check
Run the Pedestal Image Check periodically to verify the instrument's column sensor which
monitors for possible errors such as an empty column or bubbles in a sample. The Pedestal
Image Check can be used for routine quality control purposes. It also provides important
diagnostic information if a detection system component fails.
Supplies needed
To run pedestal image check
To interpret pedestal image check results
If the Pedestal Image Check displays a yellow triangle instead of the green check mark, follow
the on-screen instructions to fix any possible problems. Then rerun the Pedestal Image Check.
If the instrument fails again,
Related Topics
Thermo Scientific
• lint-free laboratory wipes
1. Lift the instrument arm and clean the upper and lower pedestal with a new laboratory
wipe.
2. Lower the arm.
3. From the instrument home screen, tap
Pedestal Image Check.
4. Tap Measure.
The instrument runs a series of tests to check pedestal position and image quality. After
the measurements are completed, the results are displayed. A green check mark indicates
the instrument passed the Pedestal Image Check.
5. When finished, tap End Experiment.
• Performance Verification
• Intensity Check
(Diagnostics) and then tap
contact
us.
19
Maintenance
Instrument Diagnostics
NanoDrop One User Guide