Current Output (Prog.) Calibration - Chroma 62020H-150S Operating & Programming Manual

Programmable dc power supply with solar array simulation
Table of Contents

Advertisement

11. Once the 7.5A point is calibrated, it is important to turn the loading of the Electronic
Load off and reconnect the DC Power Supply to a current shunt whose rating is closest to
but still covers 20A. For the Prodigit 7550, use 20A shunt directly.
12. Press "
" to perform the 19.5A calibration. The cursor will stop at position [4] as
shown in Figure 3-115 for setting the loading current of the Electronic LOAD to 19.500A.
Enter the current read by the Current Shunt (DVM) and press "
wait for it to end. Use 7.5A and 19.5A for calibration; the system will calculate the
calibration factor for the high current range.
S E T
L O A D I N G
A C T U A L O U T P U T
13. The current calibration is done once the above actions are completed. To save the
calibration data, press "
3-116. Press "
save it. If there is no need to save it, press "
EXIT
14. Press "
" to return to the MAIN PAGE.
CAUTION
Improper shunt range selection may cause damage to the current shunt.
The calibration points may be different for other models (non
62150H-600S); perform the calibration following the instructions
displayed.
3.3.7.3

Current Output (PROG.) Calibration

3.3.7.3.1
Hardware Requirements
Table 3-8 lists the hardware requirements for current output calibration.
[ C U R R E N T
C U R R E N T
C U R R E N T = 0 . 0 0 0 0 _ A
Figure 3-115
SAVE
" to display the confirmation page as shown in Figure
1
" or turn the "Rotary" (
S A V E = N O
Figure 3-116
ENTER
M E A S
C A L I B R A T I O N ]
= 1 9 . 5 0 0
A
) to set SAVE=YES and press "
EXIT
" to return to the Calibration screen.
[ S A V E A R G U M E N T ]
Manual Operation
" to confirm and
[ 4 ]
ENTER
" to
3-65