62000H Series DC Power Supplies with Solar Array Simulation may be connected in series or
parallel when in the CV/CC MODE. The maximum voltage is 1200V if connected in series and
the maximum current is 250A if connected in parallel. When the IV curve function is enabled
in the TABLE Mode, SAS Mode or IV PROGRAM Mode, only parallel connections are
supported (no series connections).
1.
2.
62000H Series
62150H-600S
62150H-1000S
62020H-150S
3.
4.
5.
6.
7.
0
9
-
Series/Parallel cannot be used at the same time. A620028 and
A620027 models only support parallel connections.
The maximum output voltage or current is 1200V or 250A when
operating the 62000H Series with Solar Array Simulation in series or
parallel. Table 3-1 shows the values for the 62xxxH-xxxS models.
Model with
Max.
Solar Array
Devices
Simulation
2
2
2
Model types cannot be mixed when operating in Series/Parallel.
Insure the breaker capacity is sufficient and the earth grounding wire
is grounded to earth ground when connecting in series/parallel.
When more than 5 devices are to be paralleled, contact the Service
Center or CHROMA agent for connection instructions.
Stack the DC Power Supplies vertically for parallel connection. The
standard CURRENT SHARING cable cannot be used if the devices
are placed horizontally in the parallel configuration; an optional
CURRENT SHARING (100CM) cable is required. Contact the
CHROMA Service Center or local agent for further information. No
more than 2 units may be paralleled when placed horizontally. .
Firmware versions 2.00 and above are not backward compatible.
Be sure to upgrade the firmware version from 1.XX to 2.00 when
connecting to a device with firmware version 2.00 for series use.
Contact Chroma's service unit for a firmware upgrade.
) or the "Rotary" (
Table 3-1
Serial
Max. Output
Voltage (V)
1200
1200
250
Manual Operation
) knob to set the range of
Parallel
Max.
Max. Output
Devices
Current (A)
10
250
10
150
10
400
3-21

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents