Programmable DC Power Supply (with Solar Array Simulation) 62000H Series
Operating & Programming Manual
S E T T I N G
A C T U A L
8.
Now start the high current range calibration. Select a current shunt whose rating is
closest to but still covers 20A. For the Prodigit 7550, use the 20A shunt
9.
Press "
" to do the 7.5A calibration. The system will set the output current to
7.500A automatically and the cursor will stop at position [3] as shown in Figure 3-122.
Input the current read by the Current Shunt (DVM) and press "
S E T T I N G
A C T U A L
10. Turn OFF the output of the DC Power Supply. Select a current shunt whose rating is
closest to but still covers 20A. For the Prodigit 7550, use the 20A shunt directly.
11. Press "
" to do the 19.5A calibration. The system will set the output current to
19.500A automatically and the cursor will stop at position [4] as shown in Figure 3-123.
Input the current read by the Current Shunt (DVM) and press "
S E T T I N G
A C T U A L
12. The current calibration is done once the above actions are completed and the DC Power
Supply output is turned OFF. To save the calibration data, pressing "
confirmation page as shown in Figure 3-124. Press "
3-68
[ C U R R E N T
O U T P U T
C U R R E N T = 4 . 5 0 0
O U T P U T
C U R R E N T = 0 . 0 0 0 _
Figure 3-121
[ C U R R E N T
O U T P U T
C U R R E N T = 7 . 5 0 0
O U T P U T
C U R R E N T = 0 . 0 0 0 _
Figure 3-122
[ C U R R E N T
O U T P U T
C U R R E N T = 1 9 . 5 0 0
O U T P U T
C U R R E N T = 0 . 0 0 0 _
Figure 3-123
S E T T I N G
C A L I B R A T I O N ]
S E T T I N G
C A L I B R A T I O N ]
S E T T I N G
C A L I B R A T I O N ]
1
" or turn the "Rotary" (
A
[ 2 ]
A
directly.
ENTER
" to confirm.
A
[ 3 ]
A
ENTER
" to confirm.
A
[ 4 ]
A
SAVE
" will display a
) to