Chroma 62020H-150S Operating & Programming Manual page 115

Programmable dc power supply with solar array simulation
Table of Contents

Advertisement

S H O R T O U T P U T
S E T T I N G
A C T U A L
4.
Turn OFF the output of the DC Power Supply before pressing ENTER. Set the Electronic
Load to CV mode 48V. Select a current shunt whose rating is closest to but still covers
2A. For the Prodigit 7550, use the 2A shunt directly.
S H O R T O U T P U T
S E T T I N G
A C T U A L
5.
Press ENTER and the system will set the output current to 1.500A automatically and the
cursor will stop at position [1] as shown in Figure 3-120. Input the current read by the
Current Shunt (DVM) and press "
S H O R T O U T P U T
S E T T I N G
A C T U A L
6.
Turn OFF the output of the DC Power Supply. A message will display to remind the user
to reconnect a current shunt. Select a current shunt whose rating is closest to but still
covers 20A. For the Prodigit 7550, use the 20A shunt directly.
7.
Press "
" to do the 4.5A calibration. The system will set the output current to
4.500A automatically and the cursor will stop at position [2] as shown in Figure 3-121.
Input the current read by the Current Shunt (DVM) and press "
1.5A and 4.5A for calibration, the system will calculate the calibration factor for the low
current range.
[ C U R R E N T
T E R M I N A L
O U T P U T
C U R R E N T = 1 . 5 0 0
O U T P U T
C U R R E N T = 0 . 0 0 0
Figure 3-118
[ C U R R E N T
T E R M I N A L
O U T P U T
C U R R E N T = 1 . 5 0 0
O U T P U T
C U R R E N T = 0 . 0 0 0
Figure 3-119
ENTER
[ C U R R E N T
T E R M I N A L
O U T P U T
C U R R E N T = 1 . 5 0 0
O U T P U T
C U R R E N T = 0 . 0 0 0
Figure 3-120
S E T T I N G
C A L I B R A T I O N ]
A N D
P R E S S [ E N T E R ]
S E T T I N G
C A L I B R A T I O N ]
A N D
P R E S S [ E N T E R ]
" to confirm.
S E T T I N G
C A L I B R A T I O N ]
A N D
P R E S S [ E N T E R ]
Manual Operation
A
A
A
A
A
A
ENTER
" to confirm. Using
3-67