Current Output (Prog.) Calibration - Chroma 62000P Series Operating & Programming Manual

Programmable dc power supply
Table of Contents

Advertisement

13. Press "
" to return to MAIN PAGE.
EXIT
Improper shunt range selection may cause damage to the current shunt.
NOTICE
The calibration point may be different for other models (non 62012P-80-60), please operate it
following the instructions displayed.

3.3.7.3 Current Output (PROG.) Calibration

3.3.7.3.1 Hardware Requirements
DVM
CURRENT SHUNT
3.3.7.3.2 SETUP
62000P
DC Power
Supply
3.3.7.3.3 Calibration Procedure (Example: Model 62012P-80-60)
(1) In CALIBRATION page, press "
ENTER
(2) Press "
to confirm and entering into current calibration options as Figure 3-86
shows.
Device
HP 34401A or equivalent DVM
Prodigit 7530 or equivalent
Table 3-10
+
sv
+
v
v
sv
Figure 3-85
" or turn "Rotary" (
3
Suggest Model or Capacity
Current Shunt
) knob to set CHOICE=3.
Manual Operation
+
DVM
3-53