Heidolph Hei-VAP Expert Operating Manual page 81

Rotary evaporators
Hide thumbs Also See for Hei-VAP Expert:
Table of Contents

Advertisement

Available languages

Available languages

Assembly
Flask immersion depth (height stop)
The stop band of the height stop is located on the right side of the lift column that can be
used to adjust a constant immersion depth for the evaporating flask.
✓ The rotation is switched off.
✓ The evaporating flask is mounted (see
chapter "Operation", section "Mounting
the evaporating flask").
✓ The height stop is at the top position.
Adjusting the lift stop without the
safety hood:
➜ Lower the lift until the desired
immersion depth is reached (see
chapter "Start-Up", section "Heating
bath and flask lift").
➜ Turn the locking screw on the stop
band of the height stop to the right
of the lift column by about 180°
counterclockwise.
➜ Slide the stop band down as far as it
will go using the locking screw.
➜ Tighten the locking screw clockwise.
➜ Return the lift to the top.
✓ The locking screw holds the stop in
position.
✓ The lift moves down no further than
the stop and ensures a constant
immersion depth of the evaporating
flask
Adjusting the lift stop with the safety
hood:
➜ Hold the safety hood with one hand.
➜ Guide the safety hood by hand when
moving the safety hood
(see chapter "Assembly, Safety hood").
Lift stop
81
Locking screw
Stop band