Operation
▪
Favorites mode (selected as a standard)
▪
Ramp profile mode
▪
Automatic mode DAA
➜ In the menu bar, tap the symbol of the
operating mode used last.
✓ The symbols of the three operating
modes are displayed.
➜ Tap the symbol of the desired operating
mode.
To switch applications, the current processes have to be stopped.
Detailed descriptions are found in the following chapters:
▪
"Grundparameter manuell einstellen („Basis")" on page 32
▪
"Favoriten verwenden und verwalten („Erweitert")" on page 36
▪
"" on page 39
▪
"Automatikmodus DAA verwenden („Erweitert")" on page 45.
Any chiller attached can be operated in all applications and modes.
Starting / stopping all processes
It is not only possible to start individual functions but also to start all functions at once.
✓ The desired settings have been made, parameters, favorites or ramp profiles have
been selected.
✓ Ensure that no glass breakage will occur.
➜ In the menu bar of the current appli-
cation, tap "Start all" (or "Start profile",
"Start DAA").
✓ The functions rotation, heating bath, pump (vacuum) and, if needed, stopwatch are
started.
✓ If a chiller is attached, it will not be started.
✓ The lift will move down all the way if it is activated in the settings.
✓ With "Start DAA", the AUTOaccurate temperature is determined prior to starting the
processes.
➜ In the menu bar of the current appli-
cation, tap "Stop all" (or "Stop profile",
"Stop DAA").
✓ All started processes are stopped.
✓ The possibly attached chiller is not stopped.
✓ The stopwatch stops and the value is displayed until the stopwatch is restarted with
"Start all".
Favorites
36
Ramps
DAA