Troubleshooting
Analyzing the list of errors
You can view a compilation of the following warnings and errors, which appear on the
display during operation, at a later time in order to analyze your process.
All in all, a maximum of 32 errors is stored. After that, older messages are overwritten by
newer ones.
➜ In the menu bar of the "Basic settings" menu, tap "Error list" to view information,
errors and warnings.
✓ An error history is displayed showing the corresponding date and time.
➜ To view details, tap the individual entries.
The following messages/errors are displayed.
Error messages on display
Warning:
Transportation lock activated!
Please move lift upward and switch
the device off and then on again.
Warning:
The vapor temperature sensor is
defective or not attached!
Warning:
The AUTOaccurate sensor is defec-
tive or not attached!
Warning from chiller:
Temperature has fallen below the
lower limit!
Warning from chiller:
Temperature has exceeded the
upper limit!
Warning from chiller:
Filling level of thermal fluid too low!
Warning:
Heating bath
Careful! Hot!
Error
The AUTOaccurate sensor is defec-
tive or the temperature is below the
measuring range.
Error
The AUTOaccurate sensor is defec-
tive or the temperature is above the
measuring range.
Error
The vapor temperature sensor is
defective or the temperature is
below the measuring range.
Cause
The transportation lock
is still activated.
Sensor is not attached
or defective
Sensor not attached or
defective
Chiller temperature too
low
Chiller temperature too
high
Not enough cooling
fluid
Heating bath tempera-
ture
> 50 °C
Defective sensor or
temperature below
-40 °C
Defective sensor or
temperature above 173
°C
Defective sensor or
temperature below
-40 °C
62
Remedy
▪ Remove transpor-
tation lock if needed
▪ Move lift upward
▪ Switch the device off
and then on again
Check vapor tempera-
ture sensor and replace
it if needed
Check AUTOaccurate
sensor and replace it if
needed
Increase chiller temper-
ature
Reduce chiller tempera-
ture
Add cooling fluid
Let the heating bath
cool down
▪ Check AUTOaccurate
sensor and replace it
if needed
▪ or increase temper-
ature
▪ Check AUTOaccurate
sensor and replace it
if needed
▪ or reduce temper-
ature
▪ Check vapor temper-
ature sensor and
replace it if needed
▪ or increase temper-
ature