:SEARch:SERial:CAN:SYMBolic:MESSage
Command Syntax
:SEARch:SERial:CAN:SYMBolic:MESSage <name>
<name> ::= quoted ASCII string
The :SEARch:SERial:CAN:SYMBolic:MESSage command specifies the message to
search for when CAN symbolic data has been loaded (recalled) into the
oscilloscope and the CAN serial search mode is set to MESSage or MSIGnal.
Query Syntax
:SEARch:SERial:CAN:SYMBolic:MESSage?
The :SEARch:SERial:CAN:SYMBolic:MESSage? query returns the specified
message.
Return Format
<name><NL>
<name> ::= quotes ASCII string
See Also
•
•
•
•
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
(see
page
1610)
":RECall:DBC[:STARt]"
":SEARch:SERial:CAN:MODE"
":SEARch:SERial:CAN:SYMBolic:SIGNal"
":SEARch:SERial:CAN:SYMBolic:VALue"
on page 813
on page 1187
on page 1194
on page 1195