Summary of Contents for Keysight Technologies InfiniiVision 6000L Series
Page 1
Keysight InfiniiVision 6000L Series Oscilloscopes User's Guide...
Page 2
No additional government performance of this document or of any requirements beyond those set forth in the information contained herein. Should Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 3
Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Connecting to the Triggering Device Under Test Displaying Measurements Printing and and Math Saving Functions Navigate to a topic Acquisition Web Interface Serial Modes Decode/Lister Mask Test Controls and Digital Channels Connectors Reference Index Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
In This User’s Guide… This guide shows you how to use the InfiniiVision 6000L Series oscilloscopes. It contains the following chapters and topics: Introduction Model numbers, options, where to find specifications. Initial Setup Unpacking and setting up your oscilloscope. Using the Quick Help system. Using the analog channels, setting up the timebase.
Page 6
Using the digital channels of a mixed-signal oscilloscope (MSO). Reference Software updates, licenses, Secure Environment mode, Measurement Category, Environmental Conditions, synchronizing the timebases of multiple instruments, cleaning, binary and csv data files, warranty status, Contact Us. Index Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Contents Book Map / 4 In This User’s Guide… / 5 Introduction Models Covered in this Manual / 21 Environmental Conditions / 22 Specifications and Characteristics / 23 Licensed Options / 24 Notes / 26 Initial Setup Package Contents / 28 6000L Series Oscilloscope Package Contents / 29 Probes and Accessories / 30 To mount the oscilloscope in a rack / 31...
Page 8
Quick Help Languages / 40 Graphical User Interface Languages / 40 Quick Help Updates / 40 To set the clock / 41 To set up the screen saver / 42 Keys, Softkeys, and the Entry Knob / 44 Using the Analog Channels / 45 To set up the Horizontal time base / 50 Connecting to the Device Under Test Measurement Category / 58...
Page 9
Triggering Triggering - General Information / 73 Trigger Mode and Coupling Menu / 74 Auto and Normal Trigger modes / 75 Choosing Auto Trigger Mode or Normal Trigger Mode / 75 Auto Mode / 75 Normal Mode / 76 Trigger Level Adjustment / 77 Trigger Coupling / 78 Trigger Noise Rejection / 79 HF Reject / 79...
Page 10
> qualifier time set softkey / 91 Edge Trigger / 93 FlexRay Setup and Triggering / 95 Setup for FlexRay Signals / 95 FlexRay Triggering / 96 Glitch or Narrow Pulse Capture / 100 Using Peak Detect Mode to Find a Glitch / 101 I2C Trigger / 103 I2S Trigger / 109 LIN Trigger / 116...
Page 11
Resetting All Bits in the Serial Data String to One Value / 144 TV Trigger / 145 TV Triggering Examples / 148 To trigger on a specific line of video / 149 To trigger on all sync pulses / 150 To trigger on a specific field of the video signal / 151 To trigger on all fields of the video signal / 152 To trigger on odd or even fields / 153...
Page 12
Vectors (Connect the Dots) / 182 Freeze Display / 183 Antialiasing / 184 XGA Video Output / 185 Measurements and Math Functions List of Automatic Measurements / 188 Making Automatic Measurements / 190 Measurement Statistics / 191 Measurement Thresholds / 193 Measurement Definitions / 196 Time Measurements / 196 Delay and Phase Measurements / 200...
Page 13
Square Root / 233 FFT Measurement / 235 FFT Operation / 237 Enabling Precision Measurements and Math / 241 Printing and Saving Printing the Oscilloscope’s Display / 244 Quick Print / 244 To print the oscilloscope’s display / 245 Options / 245 Palette / 246 Saving Oscilloscope Data / 247 Selecting a Destination for Your Saved Data...
Page 14
Peak Detect Mode / 266 High Resolution Mode / 266 Averaging Mode / 267 Realtime Sampling Option / 269 Segmented Memory / 271 Using segmented memory / 271 Web Interface Setting up the Oscilloscope’s LAN Connection / 277 To establish a LAN connection / 277 To establish a LAN connection (6000L Series) / 278 Stand-alone (Point-to-Point) Connection to a PC / 279 Accessing the Web Interface / 281...
Page 15
FlexRay Serial Decode / 305 Interpreting FlexRay Frame Decode / 306 FlexRay Totalizer / 307 Interpreting FlexRay Lister Data / 308 C Serial Decode / 309 Interpreting I C Decode / 311 Interpreting I C Lister Data / 312 S Serial Decode / 313 Interpreting I S Decode / 315 Interpreting I...
Page 16
Source Lock / 345 Mask Test Trigger Output / 346 Mask Statistics / 347 Reset Statistics / 348 Transparent / 348 To manually modify a mask file / 349 Building a Mask File / 353 Controls and Connectors 6000L Series Front and Rear Panel / 358 Front Panel Control and Connector Descriptions / 359 Rear Panel Control and Connector Descriptions / 361 Digital Channels...
Page 17
To display digital channels as a bus / 378 Digital channel signal fidelity: Probe impedance and grounding / 382 Input Impedance / 382 Probe Grounding / 384 Best Probing Practices / 385 To replace digital probe leads / 386 Utilities To restore the oscilloscope to its default configuration / 388 To perform service functions / 389 User Calibration / 390...
Page 18
Binary Data in MATLAB / 407 Binary Header Format / 407 Example Program for Reading Binary Data / 410 Examples of Binary Files / 411 CSV and ASCII XY files / 414 Acknowledgements / 415 To check warranty and extended services status / 416 To return the instrument / 417 Contact Us / 418 Index...
6000 Series models. • 4-channel Digital Storage Oscilloscope (DSO) models. • You can easily upgrade an InfiniiVision 6000L Series oscilloscope from a DSO to an MSO and/or add features such as Segmented Memory and Mask Test. An MSO lets you debug your mixed-signal designs using up to four analog signals and 16 tightly correlated digital signals simultaneously.
Page 20
Introduction For more information about InfiniiVision oscilloscopes, see: www.keysight.com/find/scope Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Occasionally a temporary conductivity caused by condensation may occur. Example: General indoor environment. Pollution Degree 3: Conductive pollution occurs, or dry, non-conductive pollution occurs which becomes conductive due to condensation which is expected. Example: Sheltered outdoor environment. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
6000L series oscilloscopes data sheet. To order a data sheet by phone, please contact your local Keysight office. The complete list is available at: www.keysight.com/find/contactus or on page page 418. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Order U7244A. Application runs on a PC Test Application connected to InfiniiVision oscilloscope. Mask Limit Test Order N5455A after purchase (Option LMT at time of purchase). You can easily install this option yourself. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 25
S serial decode option (for 4 channel of purchase). You can easily install this option or 4+16 channel models only) yourself. See also “Installed Licenses" on page 398. Calibration Option Option Order ANSI Z540 Compliant Calibration Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
6000L Series oscilloscope, and this book refers to using front panel controls, you can use the built-in Web control feature described on “Accessing the Web Interface" on page 281 to complete the instructions. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 27
Keysight InfiniiVision 6000L Series Oscilloscopes User's Guide 2 Initial Setup Package Contents / 28 6000L Series Oscilloscope Package Contents / 29 Probes and Accessories / 30 To mount the oscilloscope in a rack / 31 Ventilation Requirements / 35 Power Requirements / 36...
P/N 0960-0301, Qty. 4 GPIB cable extender P/N 5183-0803 Oscilloscope Probes 10073C or 10074C Qty. 4 Rack Mount Kit (not shown) Documentation CD Automation-Ready CD Power cord (Based on country of origin) Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
InfiniiVision Oscilloscope Probes and Accessories Selection Guide Data Sheet (5968-8153EN) • For compatibility information, manuals, application notes, data sheets, selection guides, SPICE models, and more for oscilloscope probes, see the Probe Resource Center at: www.keysight.com/find/PRC Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 32
2 Fasten the Rack Mount Extenders to the oscilloscope chassis with the four (4) M3 x 6 mm screws, using a Torx T10 driver as follows: Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 33
(4) remaining 10-32 x 0.375 Rail Screws. Use the Torx T20 driver. 5 Securely attach the Rear Extender Supports to the Front Extender Supports by tightening the four (4) 10-32 x 0.375 Rail Screws screws that you loosely attached in step Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 34
Initial Setup 6000L Mounting Bracket Dimensions The following information is provided in case you want to design custom mounting brackets for the 6000L. Dimensions are in mm. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
6000L Series Ventilation Requirements The fan draws air from the left and pushes it out to the right. Ensure the rack in which the oscilloscope is mounted is properly ventilated. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Initial Setup Power Requirements 6000L Models Line voltage, frequency, and power ~Line 100-240 Vac, 50/60 Hz 80 W max Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
2 Press the power switch. The power switch is located on the lower left corner of the front panel. The oscilloscope will perform a self-test and will be operational in a few seconds. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
If you do not see the waveform, ensure your power source is adequate, the oscilloscope is powered-on, the power cord is firmly inserted into the oscilloscope receptacle, and the probe is connected securely to the front-panel oscilloscope channel input BNC and to the Probe Comp terminal. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Start a Web browser control session as described on “Accessing the Web Interface" on page 281 and select Remote Front Panel. To view Quick Help information, right-click on the softkey. Help is not available for front panel keys; only softkeys. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Updated Quick Help may be made available. To download new Quick Help and GUI language files: 1 Direct your web browser to: www.keysight.com/find/6000sw (for 6000 models) 2 Select Quick Help Language Support and follow the directions. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
The hours are shown in the 24-hour format. So 1:00 PM is hour 13. The real-time clock only allows selection of valid dates. If a day is selected and the month or year is changed so the day is invalid, the day is automatically adjusted. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
3 Turn the Entry knob to select the number of minutes to wait before the selected screen saver activates. When you turn the Entry knob, the number of minutes is displayed in the Wait softkey. The default time is 180 minutes (3 hours). Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 43
4 Press the Preview softkey to preview the screen saver you have selected with the Saver softkey. 5 To view the normal display after the screen saver has started, press any key or turn any knob. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
The simplest way to set up the oscilloscope is to connect it to the signals of interest and press NOTE the [AutoScale] key. To find out how to use a key or connector, see “Controls and Connectors" page 357. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Pressing an analog channel key displays the channel’s menu and turns the display of the channel on or off. The channel is displayed when the key is illuminated. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 46
( ) icon. It also represents the voltage at the vertical center of the display if vertical expansion is set to expand about ground. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 47
When you connect an AutoProbe, self-sensing probe, or a compatible InfiniiMax probe, the NOTE oscilloscope will automatically configure the oscilloscope to the correct impedance. The analog channel input impedance can be set to either 1M Ohm or 50 Ohm by pressing the Imped softkey. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 48
Inverting a channel will also change the result of any function selected in the Waveform Math Menu or any measurement. 7 Press the Probe softkey to display the Channel Probe Menu. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 49
±100 ns in 10 ps increments for a total of 200 ns difference. The skew setting is not affected by pressing [Save/Recall]&Default Setup or by pressing [Auto Scale]. • Calibrate Probe — See “Calibrating Probes" on page 67. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Horizontal knobs to pan and zoom the waveform. 2 Turn the large knob (time/division) in the Horizontal section and notice the change it makes to the status line. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 51
Changing this delay value does not effect the position of the normal window. 5 Turn the delay knob ( ) and notice that its value is displayed in the status line. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 52
The area of the normal display that is expanded is outlined with a box and the rest of the normal display is lightened. The box shows the portion of the normal sweep that is expanded in the lower half. The Horizontal Sweep Speed controls the size Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 53
Zoom window sweep Horizontal Position knob Zoom window is turned Normal sweep window Signal anomaly expanded in zoom window Zoom window Select Normal sweep or Zoom Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 54
Events that have occurred are scrolled to the left of the reference point. Since there is no trigger, no pre-trigger information is available. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 55
Knob softkey The “Knob “softkey is used in conjunction with Segmented Memory. You can set the Horizontal Position knob to adjust the horizontal position or move between segments. See “Segmented Memory" on page 271. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Measurement category IV is for measurements performed at the source of the low-voltage installation. Examples are electricity meters and measurements on primary overcurrent protection devices and ripple control units. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Impedance Selection" on page 47. For more information on probing, visit www.keysight.com/find/scope_probes Information about selecting a probe can be found in document number 5989-6162EN: Keysight Oscilloscope Probes and Accessories Selection Guide, available at www.keysight.com. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
BNC connector to transfer information between the oscilloscope and the probe. When you connect a compatible probe to the oscilloscope, the AutoProbe interface determines the type of probe and sets the oscilloscope’s parameters (units, offset, attenuation, coupling, and impedance) accordingly. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
1158A 4 GHz Active Probe N2772A Differential Probe, 20 MHz, with N2773A power supply N2774A (obsolete, replaced by N2782A) with N2775A power supply N2782A 50 MHz/30Arms AC/DC Current Probe, with N2779A power supply Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Active Probes for All 6000 Series Oscilloscopes Active Probes Quantity Supported 1141A with 1142A power supply 1144A with 1142A power supply 1145A with 1142A power supply N2772A with N2773A power supply N2774A with N2775A power supply Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
If you need to measure between two live points, use a differential probe with sufficient dynamic range. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 65
Connecting to the Device Under Test Do not negate the protective action of the ground connection to the oscilloscope. The WARNING oscilloscope must remain grounded through its power cord. Defeating the ground creates an electric shock hazard. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
BNC connector. Perfectly compensated Over compensated Under compensated comp.cdr 3 Connect probes to all other oscilloscope channels (channels 2, 3, and 4 of a 4-channel oscilloscope). Repeat the procedure for each channel. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
10:1 differential browser (no attenuator) • 10:1 (+6 dB Atten) single-ended browser • 10:1 (+6 dB Atten) differential browser • 10:1 (+12 dB Atten) single-ended browser • 10:1 (+12 dB Atten) differential browser Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 68
• 10:1 (+20 dB Atten) differential browser 6 Press the Calibrate Probe softkey and follow the instructions on the display. For more information on InfiniiMax probes and accessories, see the probe’s User’s Guide. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
If Amps is chosen as the units and a manual attenuation factor is chosen, then the units as well as the attenuation factor are displayed above the Probe softkey. Units and Attenuation Factor Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Connecting to the Device Under Test Digital Probes Please see “To connect the digital probes to the device under test" on page 364 for information about digital probing. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Keysight InfiniiVision 6000L Series Oscilloscopes User's Guide 4 Triggering Triggering - General Information / 73 Trigger Mode and Coupling Menu / 74 Auto and Normal Trigger modes / 75 Trigger Level Adjustment / 77 Trigger Coupling / 78 Trigger Noise Rejection / 79...
Page 72
[AutoScale] key and the oscilloscope will attempt to trigger on the waveform using a simple Edge Trigger type. For AutoScale details see “AutoScale" on page 168. Complex waveforms may require specific trigger setups as described in this chapter. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
(I C and SPI, or CAN and LIN). Adding option FRS (FlexRay serial decode) will add the FlexRay triggers. • See also: “Segmented Memory" on page 271 Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Triggering Trigger Mode and Coupling Menu • Press the [Mode/Coupling] key in the Trigger section of the front panel. 2-channel models only Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
(FIFO). When a trigger is found, the pre-trigger buffer will contain the events that occurred just before the trigger. If no Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Some measurements you want to make will require you to take some action in the device under test to cause the trigger event. Usually, these are single-shot acquisitions, where you will use the [Single] key. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
The threshold value is displayed in the upper-right corner of the display. The line trigger level is not adjustable. This trigger is synchronized with the power line supplied to the oscilloscope. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
TV coupling is normally grayed-out, but is automatically selected when TV trigger is enabled in the Trigger More Menu. Note that Trigger Coupling is independent of Channel Coupling. To change Channel Coupling see page Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
50 kHz LF Reject Low-frequency reject (LF reject) adds a high-pass filter with the 3-dB point at 50 kHz. LF reject removes low-frequency signals, such as power line noise, from the trigger path. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
However, this also decreases the trigger sensitivity so that a slightly larger signal is required to trigger the oscilloscope. 1 Press [Mode/Coupling]&HF Reject. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Use Holdoff to stabilize the display of complex waveforms. To get a stable trigger on the pulse burst shown below, set the holdoff time to be >200 ns but <600 ns. Holdoff Scope triggers here 200 ns 600 ns Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 82
The oscilloscope will then re-arm the trigger circuit to search for the next edge trigger. This allows the oscilloscope to trigger on a repeating pattern in a waveform. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
15 Vrms. Do not exceed 15 Vrms at rear panel External Trigger input or damage to the oscilloscope may occur. There are no range or units settings for the External Trigger input of the 4-channel oscilloscope. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
This mode is useful when your input signal is faster than 350 MHz. The Trigger Out connector also provides the User Cal signal. See “User Calibration" on page 390. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
[Run/Stop] or [Single]. If the oscilloscope is running when you change a triggering specification, it uses the new trigger definition when it starts the next acquisition. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
4 Press the [More] key in the Trigger section of the front panel, rotate the Entry knob until CAN is displayed in the Trigger softkey. Currently selected Trigger level or signal source threshold CAN trigger Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 87
All Errors - The oscilloscope will trigger when any form error or active error is encountered. • Acknowledge Error - The oscilloscope will trigger when the acknowledge bit is recessive (high). • Overload Frame - The oscilloscope will trigger on CAN overload frames. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 88
CAN signal (CAN_L) and connect the negative lead to the dominant-high CAN signal (CAN_H). 10 Repeatedly press and release the Baud softkey to set the CAN signal baud rate to match your CAN bus signal. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 89
If the setup does not produce a stable trigger, the CAN signal may be slow enough that the NOTE oscilloscope is AutoTriggering. Press the [Mode/Coupling] key, then press the Mode softkey to set the trigger mode from Auto to Normal. To display CAN serial decode, see page 297. NOTE Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
• 0 sets the pattern to 0 (low) on the selected channel. A low is a voltage level that is less than the channel's trigger level or threshold level. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
• When the within time range (><) qualifier is selected, the Entry knob sets the lower time range value. • When the out of time range (<>) qualifier is selected, the Entry knob sets the upper time range value. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 92
Timeout mode. In Timeout, the trigger occurs when the timeout value is reached while the pattern is true. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Either edge mode will trigger on Constant Wave signals up to 100 MHz, but can trigger on isolated pulses down to 1/(2*oscilloscope's bandwidth). 3 Select the trigger source. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 94
The selected trigger source is indicated in the upper-right corner of the display next to the slope symbol: 1 through 4 = analog channels D0 through D15 = digital channels E = External trigger L = Line trigger Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
5 In the FlexRay Signals Menu, press Source and select the analog channel that is probing the FlexRay signal. 6 Press Baud and select the baud rate of the FlexRay signal being probed. 7 Press Bus and select the bus type of the FlexRay signal being probed. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
1 In the front panel Trigger section, press [More]. 2 In the More Trigger Menu, press Trigger; then, select the FlexRay trigger type. 3 Press Settings. 4 In the FlexRay Trigger Menu, press Trigger; then, select Frame. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 97
Normal trigger mode instead of Auto trigger mode. This prevents the oscilloscope from Auto triggering while waiting for a particular frame and cycle combination. Triggering on FlexRay Errors 1 In the front panel Trigger section, press [More]. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 98
1 In the front panel Trigger section, press [More]. 2 In the More Trigger Menu, press Trigger; then, select the FlexRay trigger type. 3 Press Settings. 4 In the FlexRay Trigger Menu, press Trigger; then, select Event. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 99
BSS - ByteStart Sequence • FES/DTS - Frame End or Dynamic Trailing Sequence 6 Press Auto Setup for Event. This automatically configures oscilloscope settings (as shown on the display) for the selected event trigger. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Normal acquire mode, making them easier to see. To characterize the glitch, use the cursors or the automatic measurement capabilities of the oscilloscope. 15 ns Narrow Pulse, 20 ms/div, Normal Mode Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Waveform persistence is not kept beyond the display area boundary. Press the Clear Display softkey to erase previously acquired points. The display will accumulate points until ∞ Persist is turned off. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 102
To obtain a better resolution of the glitch, expand the time base. Use the horizontal position knob ( ) to pan through the waveform to set the expanded portion of the normal sweep around the glitch. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
SDA Data SCL Clock previous menu channel channel 6 Connect an oscilloscope channel to the SCL (serial clock) line in the device under test, then set the SCL clock channel softkey to that channel. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 104
Start Condition — The oscilloscope triggers when SDA data transitions from high to low while the SCL clock is high. For triggering purposes (including frame triggers), a restart is treated as a start condition. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 105
When this event occurs, the oscilloscope will trigger on the clock edge for the Ack bit after the data byte. This data byte does not need to occur directly after the control byte. Read Control byte Data Start or Trigger point Restart Condition Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 106
10-bit Write — The oscilloscope triggers on a 10-bit write frame on the 26th clock edge if all bits in the pattern match. The frame is in the format: Frame (Start: Address byte 1: Write: Address byte 2: Ack: Data) Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 107
The trigger will always occur on the 17th clock for 7-bit addressing or 26th clock for 10-bit addressing. b Press the Data value softkey and turn the Entry knob to select the 8-bit data pattern on which to trigger. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 108
If you have selected a three-byte trigger, press the Data2 value softkey and turn the Entry knob to select the 8-bit data pattern on which to trigger. To display I C serial decode, see page 313. NOTE Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
5 Press the Settings softkey to display the I S Trigger Menu. Currently selected channel Trigger level or threshold S trigger Return to I2S trigger Signals setup menu previous menu configuration menu menu Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 110
“I S”. 8 Connect an oscilloscope channel to the WS (word select) line in the device under test, switch the channel on, then set the WS channel softkey to that channel. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 111
WS, SCLK, and SDATA signals for the currently specified bus configuration. 12 Press the Word Size softkey. Rotate the Entry knob to match the transmitter word size of the device under test (from 4 to 32 bits). Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 112
Right-Justified: Data transmission (MSB first) is right-justified to the transition of 15 Press the WS Low softkey to select whether WS Low indicates Left or Right channel data. The on-screen diagram changes with your selection. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 113
20 Press the Trigger softkey and choose a qualifier: • Equal — triggers on the specified audio channel’s data word when it equals the specified word. • Not equal — triggers on any word other than the specified word. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 114
“Trigger“ value. If you select an “Armed” value that is equal to or greater than the “Trigger” value, the “Trigger” value will be increased so that it is always greater than the “Armed” value. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 115
AutoTriggering. Press the [Mode/Coupling] key, then press the Mode softkey to set the trigger mode from Auto to Normal. To display I S serial decode, see page 313. NOTE Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Trigger level or signal source threshold LIN trigger 4 Press the Settings softkey to display the LIN Trigger Menu. Return to Enter data Trigger Select Signals values previous menu Frame ID menu condition Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 117
User Baud softkey and rotate the Entry knob to make your selection. 8 Press the Smpl Pt softkey to select the sample point at which the oscilloscope will sample the bit value. One Bit Sample Point Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 118
Hex or Binary previous menu value nibbles/bits (up to 8) position entry For details about using the LIN Bits Menu softkeys, press and hold the softkey in question to NOTE display the built-in help. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 119
Triggering For LIN decode information see page 317. NOTE Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
If you use a different differential probe that is designed for 50 Ohm impedance oscilloscope inputs, be sure to set only one of the oscilloscope channel input impedances to 50 Ohms (because the two input channel’s impedances are seen in parallel). Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Threshold Channel trigger level to -500 mV. • Turns off trigger noise reject. • Turns on Serial Decode. • Sets the trigger type to MIL-1553. These sofkeys appear in both the Trigger Menu and the Serial Decode Menu. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
When this option is selected, the RTA + Bits softkey appears and opens the MIL-STD-1553 RTA + 11 Bits Menu. In this menu: • The RTA softkey lets you select the hex Remote Terminal Address value. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 123
Manchester Error – triggers if a Manchester encoding error is detected. • Sync Error – triggers if an invalid Sync pulse is found. To display MIL-STD-1553 serial decode, see “MIL-STD 1553 Serial Decode" on page 323. NOTE Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
1 Press the [More] key in the Trigger section of the front panel, rotate the Entry knob until Nth Edge Burst is displayed in the Trigger softkey. 2 Press the Settings softkey to display the Nth Edge Burst Trigger Menu. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 125
3 Press the Slope softkey to specify the slope of the edge. 4 Press the Idle softkey; then, turn the Entry knob to specify the idle time. 5 Press the Edge softkey; then, turn the Entry knob to which edge to trigger on. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
The value of the trigger level or digital threshold is displayed in the upper-right corner of the display. 3 For each channel you select, press one of the condition softkeys to set the condition for that channel in the pattern. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
If a digit is made up of less than four bits, then the value of the digit will be limited to the NOTE value that can be created by the selected bits. 5 You can use the Set all Digits softkey to set all digits to a particular value. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 128
0 or 1, the “$” sign will be displayed for the digit. For information regarding digital bus display when Pattern triggering see “Bus values are displayed when using Pattern trigger" on page 380. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
The selected pulse polarity is displayed in the upper-right corner of the display. A positive pulse is higher than the current trigger level or threshold and a negative pulse is lower than the current trigger level or threshold. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
< qualifier time set softkey • When the less than (<) qualifier is selected, the Entry knob sets the oscilloscope to trigger on a pulse width less than the time value displayed on the softkey. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
• When the time range (><) qualifier is selected, the Entry knob sets the lower time range value. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
To access the sequence trigger settings, press the [More] key in the Trigger section of the front panel, rotate the Entry knob until Sequence is displayed in the Trigger softkey, then press the Settings softkey to display Sequence Trigger Menu. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
As you set stage, term, and channel definitions for the sequence trigger, these settings will be shown in the waveform area of the display. Define the “Find” Stage 1 Press the Stage softkey and select Find:. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 134
Trigger Level knob. Press the [D15-D0] key and select Thresholds to set the threshold level for digital channels. The value of the trigger level or digital threshold is displayed in the upper-right corner of the display. c Repeat for all channels in the pattern. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
2 Press the Trigger: softkey and select a stage to trigger on. 3 To define the term(s) used in the Trigger on: stage, press the Term softkey and select the pattern or edge term displayed in the Trigger: softkey. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
At least one term in the stage must be set to a value other than X (don’t care). Define the Optional “Reset on” Stage 1 Press the Stage softkey and select Reset on:. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 137
Timeout can be set from 10 ns to 10 seconds. The timer starts when the Find condition has been met. If the timer is running and another Find condition occurs, the timer will restart from time zero. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
• To set the threshold level for digital channels, press the [D15-D0] key and select Thresholds. The value of the trigger level or digital threshold is displayed in the upper-right corner of the display. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
4 Press the [More] key in the Trigger section of the front panel, rotate the Entry knob until SPI is displayed in the Trigger softkey, then press the Settings softkey to display the SPI Trigger Menu. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
# data bits Assign value bits to value select previous menu in string channels Assign Oscilloscope Channels to SPI Signals This explains how to assign oscilloscope channesl to the SPI clock, data, and frame signals. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 141
This determines which clock edge the oscilloscope will use to latch the serial data. When you press the slope softkey, the graphic shown on the display changes to show the current state of the clock signal. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 142
Thresholds to set an appropriate threshold level for digital channels. 5 Press the Frame by softkey to select a framing signal that the oscilloscope will use for determining which clock edge is the first clock edge in the serial stream. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 143
The Timeout value can be set anywhere from 100 ns to 10 s. 6 Press the up-arrow softkey to return to the previous menu. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
1 Press the 0 1 X softkey to select your reset bit value. 2 Press the Set all Bits softkey to reset the data string to the selected value. For SPI decode information see page 327. NOTE Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
1 Press the [More] key in the Trigger section of the front panel. If TV is not selected, rotate the Entry knob until TV is displayed in the Trigger softkey, then press the Settings softkey to display TV Trigger Menu. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 146
75 Ω terminator (such as an Keysight 11094B) should be connected to the oscilloscope input. 3 Press the sync polarity softkey to set the TV trigger to either positive ( ) or negative ( ) sync polarity. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 147
70 μs after the start of vertical sync, whichever occurs first (only available in Generic mode). • Count: Vertical — Counts falling edges of sync pulses; triggers on the selected count number (only available in Generic mode). Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Field 1 and/or Field 2. For a non-interlaced video signal, the counting starts after the rising edge of the vertical sync pulse. TV Triggering Examples The following are exercises to familiarize you with TV triggering. These exercises use the NTSC video standard. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Field 1 and Field 2. This is a quick way to compare the Field 1 VITS and Field 2 VITS or to check for the correct insertion of the half line at the end of Field 1. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
1 Press the Trigger [More] key, then press the TV softkey. 2 Press the Settings softkey, then press the Standard softkey to select the appropriate TV standard. 3 Press the Mode softkey and select All Lines. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
1 Press the Trigger [More] key, then press the TV softkey. 2 Press the Settings softkey, then press the Standard softkey to select the appropriate TV standard. 3 Press the Mode softkey and select Field1 or Field2. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
1 Press the Trigger [More] key, then press the TV softkey. 2 Press the Settings softkey, then press the Standard softkey to select the appropriate TV standard. 3 Press the Mode softkey and select All Fields. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Line 4. In the case of NTSC video, the oscilloscope will trigger on color field 1 alternating with color field 3 (see the following figure). This setup can be used to measure the envelope of the reference burst. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 154
Repeat until the correct phase is displayed. When holdoff is adjusted using the TV Holdoff softkey and the Entry knob, the corresponding holdoff time will be displayed in the Trigger Mode and Coupling Menu. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 155
HDTV 720p/60 8.35 ms HDTV 1080p/24 20.835 ms HDTV 1080p/25 20 ms HDTV 1080i/50 10 ms HDTV 1080i/60 8.35 ms Using TV Holdoff to Synchronize to Color Field 1 or 3 (Field 1 mode) Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
5 Press the Settings softkey to display the UART/RS232 Trigger Menu. Currently Trigger level or selected Rx or threshold Tx channel UART/RS232 trigger mode Return to Trigger Trigger Define Define Setup previous menu Base Signal Channels Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 157
Bit Order: Select whether the most significant bit (MSB) or the least significant bit (LSB) is presented after the start bit in the signal from your device under test. For RS232 select LSB. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 158
Data is softkey. 17 Optional: The Burst softkey lets you trigger on the Nth frame (1-4096) after an idle time you select. All trigger conditions must be met for the trigger to occur. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 159
Press the Idle softkey and rotate the Entry knob to set an idle time. 19 Remember to set the trigger mode to Normal (press [Mode/Coupling]&Mode&Normal) to prevent Auto Triggering. To display UART/RS232 serial decode, see page 333. NOTE Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Return to condition speed channel channel previous menu 5 Press the Speed softkey to select the speed of the transaction being probed. You may select Low Speed (1.5 Mb/s) or Full Speed (12 Mb/s). Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 161
Exit Sus (Exit Suspend) — triggers when exiting an idle state > 10 ms. This is used to see the suspend/resume transition. End of packet trigger D+,– Bus Idle D–,+ SE0 portion of EOP Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 162
Triggering Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 163
Keysight InfiniiVision 6000L Series Oscilloscopes User's Guide 5 Displaying Tips for Displaying Waveforms / 164 Stabilizing the Display / 165 Interpreting the Display / 166 Graphic Symbols in Softkey Menus / 167 AutoScale / 168 Pan and Zoom / 170...
The easiest way to display waveforms is to connect the probes to the signals and press [AutoScale]. When selecting a particular trigger type, it is helpful to consider whether to use Auto or Normal trigger mode. See “Auto and Normal Trigger modes" on page 75. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
If the acquired signal is noisy, there are several steps you can take to stabilize its display. To reduce the random noise displayed when triggering on a noisy signal, “Trigger Noise Rejection" on page 79. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
This line normally contains automatic measurement and cursor results, but can also display advanced trigger setup data and menu information. Softkeys The softkeys let you set up additional parameters for the selected mode or menu. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Feature is on. Press the softkey again to turn the feature off. Feature is off. Press the softkey again to turn the feature on. Press the softkey to view the menu. Press the softkey to return to the previous menu. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
DC voltage, ground, or an active AC signal. Channel coupling is maintained for easy viewing of oscillating signals. 1 Press [Utility] > Options > Preferences > AutoScale. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Choose Normal to make the oscilloscope switch to Normal acquisition mode whenever the [AutoScale] key is pressed. This is the default mode. Choose Preserve to make the oscilloscope remain in the acquisition mode you have chosen when the [AutoScale] key is pressed. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
The [Run/Stop] key is illuminated red when the oscilloscope is stopped. 2 Turn the horizontal sweep speed knob marked to zoom horizontally and turn the volts/division knob marked to zoom vertically. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 171
( ) to pan vertically. The stopped display may contain several triggers worth of information, but only the last trigger acquisition is available for pan and zoom. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Reducing the intensity can expose more detail in complex signals as shown in the following figures. The intensity knob does not affect the digital channels. Amplitude Modulation with Noise Shown at 100% Intensity Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 173
Displaying Amplitude Modulation with Noise Shown at 40% Intensity Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Each major horizontal division in the grid corresponds to the sweep speed time shown in the status line at the top of the display. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
The display is cleared if you press the Clear Display softkey or if you press the [AutoScale] key. See also “Glitch or Narrow Pulse Capture" on page 100. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
The channel labels default to their channel number. 2 To turn the labels off, press the [Label] key until it is not illuminated. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 177
4 Press the Apply New Label softkey to assign the label to your selected channel. 5 Repeat the above procedure for each predefined label you want to assign to a channel. To define a new label 1 Press the [Label] key. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 178
You can create a list of up to 75 labels and load it into the oscilloscope. Labels are added to the beginning of the list. If more than 75 labels are loaded, only the first 75 are stored. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 179
This will delete all user-defined labels from the library and set the labels in the library back to the factory default. However, this does not default the labels currently assigned to the channels (those labels that appear in the waveform area). Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 180
Defaulting labels without erasing the default library. Selecting Default Setup in the Save/Recall Menu will set all channel labels back to the default labels but does not erase the list of user-defined labels in the library. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
If the ground level is off screen, the waveform will expand about the top or bottom edge of the screen based on where the ground is off screen. Expand About Center The displayed waveform will expand about the center of the display. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
• Turn vectors off when highly complex or multivalued waveforms are displayed. Turning vectors off may aid the display of multivalued waveforms such as eye diagrams. • Having vectors on does not slow down the display rate. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
To un-freeze the display, press Freeze Display again. Many activities, such as adjusting the trigger level, adjusting vertical or horizontal settings, or saving data will un-freeze the display. Manual cursors can be used on the frozen display. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
If you need to switch Antialiasing off, press [Utility]&Options&Preferences and press the Antialiasing softkey to switch the feature off. The displayed waveforms will be more susceptible to aliasing. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
The oscilloscope’s built-in display will remain on even when an external display is connected. For optimal video quality and performance we recommend you use a shielded video cable with ferrite cores. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 186
Displaying Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
204. • “Maximum" on page 204. • “Minimum" on page 204. • “Peak-Peak" on page 205. • “Ratio" on page 205. • “RMS" on page 205. • “Std Deviation" on page 205. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 189
Measurements and Math Functions • “Top" on page 207. Preshoot and Overshoot • “Preshoot" on page 208. • “Overshoot" on page 208. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
4 The Settings softkey will be available to make additional measurement settings on some measurements. 5 Press the Measure softkey to make the measurement. (By default, measurement statistics are displayed. See page 191.) Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
1. Select the measurements you desire for the channels you are using (see page 188). From the Measurement Menu, press the Statistics softkey to enter the Statistics Menu. Statistics Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 192
When you navigate away from the Measurements Menu, the statistics display will no longer be displayed, but statistical data collection will continue. Return to the Measurements Menu to see the data again. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Top and Base. Changing these threshold definitions from the default values may change the returned measurement results for Average, Delay, Duty Cycle, Fall Time, Frequency, Overshoot, Period, Phase, Preshoot, Rise Time, +Width, and -Width. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 194
If threshold Type is set to %, the lower threshold value can be set from 5% to 93%. 5 Press the Middle softkey, then turn the Entry knob to set the middle measurement threshold value. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 195
The default upper threshold is 90% or 1.50 V. • If threshold Type is set to %, the upper threshold value can be set from 7% to 95%. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
100 ms or twice the current time window, whichever is longer, up to 1 second. The Counter can measure frequencies up to the bandwidth of the oscilloscope. The minimum frequency supported is 1/(2 X gate time). Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 197
Zoom mode to isolate an event for a frequency measurement. If the measurement is not possible in the Zoom mode, then the normal time base is used. If the waveform is clipped, it may not be possible to make the measurement. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 198
The rise time of a signal is the time difference between the crossing of the lower threshold and the crossing of the upper threshold for a positive-going edge. The X cursor shows the edge being measured. For maximum measurement accuracy, set Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 199
Minimum, starting from the left-side of the display. For periodic signals, the position of the minimum may vary throughout the waveform. The X cursor shows where the current X at Min Y value is being measured. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
The default Delay settings measure from the rising edge of channel 1 to the rising edge of channel 2. 2 Press the Measure Delay softkey to make the measurement. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 201
Source 1 Delay Source 2 1 Press the Settings softkey to select the source 1 and source 2 channels for the phase measurement. The default Phase settings measure from channel 1 to channel 2. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
1-2 and for d/dt, and dt when 1-2 or 1+2 is the selected source if channel 1 and channel 2 are set to dissimilar units in the channel Probe Units softkey. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 203
Automatic voltage measurements are not valid on digital channels on the mixed-signal oscilloscope. Amplitude The Amplitude of a waveform is the difference between its Top and Base values. The Y cursors show the values being measured. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 204
Maximum is the highest value in the waveform display. The Y cursor shows the value being measured. Minimum Minimum is the lowest value in the waveform display. The Y cursor shows the value being measured. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 205
It is useful, for example, for measuring power supply noise. The standard deviation of a measurement is the amount that a measurement varies from the mean value. The Mean value of a measurement is the statistical average of the measurement. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 206
------------- - where: x = the mean. N = the number of measurements taken. = the i measurement result. The standard deviation is calculated as follows: – σ ----------------------------- - – where: Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Isolating area for Top measurement Overshoot and Preshoot Measurements Digital channel time measurements Automatic measurements Preshoot and Overshoot are not valid measurements for the math FFT function or for digital channels on the mixed-signal oscilloscope. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 208
Maximum - D Top Rising edge overshoot = --------------------------------------------------------- - X 100 Amplitude Base - D local Minimum Falling edge overshoot = ---------------------------------------------------------- - X 100 Amplitude Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 209
Because measurements and math functions are performed on displayed data, you affect the resolution of functions and measurements. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
The cursor values for the selected waveform source are displayed within the Y1 and Y2 softkeys. The difference between Y1 and Y2 (ΔY) is displayed on the dedicated line above the softkeys or in the display area when some menus are selected. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
The display is color coded to match the color of the related channel’s waveform. Cursor X1 values for Cursor X1 values for analog channels 1 digital channels through 4 D15 Through D0 Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 212
Y1 Y2 — adjusts both cursors at the same time, while the delta value remains the same. Then, turn the Entry knob. The cursor currently assigned to the Entry knob displays brighter than the other cursors. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
After adjusting the X1, X2, X1 X2, Y1, Y2, or Y1 Y2 controls, you can press the Entry knob to NOTE adjust the next control. Cursor Examples Cursors used to measure pulse widths other than middle threshold points Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 214
Measurements and Math Functions Cursors measure frequency of pulse ringing Expand the display with Zoom mode, then characterize the event of interest with the cursors. Cursors track Zoom sweep Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 215
Put the X1 cursor on one side of a pulse and the X2 cursor on the other side of the pulse. Measuring pulse width with cursors Press the X1 X2 softkey and move the cursors together to check for pulse width variations in a pulse train. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 216
Measurements and Math Functions Moving the cursors together to check pulse width variations Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
The phase difference angle (θ) can be calculated using the following formula (assuming the amplitude is the same on both channels): θ -- - or --- - Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 218
5 Set the Y2 cursor to the top of the signal, and set Y1 to the bottom of the signal. Note the ΔY value at the bottom of the display. In this example, we are using the Y cursors, but you could have used the X cursors instead. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 219
6 Move the Y1 and Y2 cursors to the intersection of the signal and the Y axis. Again, note the ΔY value. Cursors set to center of signal 7 Calculate the phase difference using the formula below. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 220
When Z is low (<1.4 V), Y versus X is displayed; when Z is high (>1.4 V), the trace is turned off. Signals are 90 out of phase Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 221
Measurements and Math Functions Signals are in phase Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
To perform a math function upon an arithmetic operation To perform a function (differentiate, integrate, FFT, or square root) on an arithmetic operation (add, subtract, or multiply): 1 Press the Function softkey and select g(t): Internal. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Units Units for each input channel can be set to Volts or Amps using the Units softkey in the channel’s Probe Menu. Scale and offset units are: Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 224
A scale unit of U (undefined) will be displayed for math functions when two source channels are used and they are set to dissimilar units and the combination of units cannot be resolved. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
(Amps-squared), or W (Watts) and is represented by the center horizontal grid line of the display. Press the Offset softkey, then turn the Entry knob to change the offset for the math waveform. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 226
Measurements and Math Functions Example of Multiply Channel 1 Channel 2 1 * 2 waveform Displayed Multiply First Second Math Math operator operand operand waveform waveform function selected scale offset Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
A scale unit of U (undefined) will be displayed for scale and offset if the two source channels are set to dissimilar units. Units are set using the channel’s Probe Units softkey. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 228
Measurements and Math Functions Example of Subtract Channel 1 Channel 2 1 – 2 waveform Subtraction First Second Math Math Displayed operand function operator operand waveform waveform selected scale offset Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Offset softkey, then turn the Entry knob to change the offset for d/dt. A scale unit of U (undefined) will be displayed for scale and offset if the two source channels are set to dissimilar units. Units are set using the channel’s Probe Units softkey. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 230
Measurements and Math Functions Example of Differentiate Function d/dt waveform Channel 1 Displayed Differentiate Source Math Math operator waveform waveform function selected scale offset Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Offset softkey, then turn the Entry knob to change the offset for dt. The integrate calculation is relative to the source signal’s offset. The following examples illustrate the effects of signal offset. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 232
Measurements and Math Functions Integrate and Signal Offset Channel 1 dt waveform Channel 1 dt waveform Displayed Integrate Source Math Math operator waveform waveform function selected scale offset Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
√ (square root). A scale unit of U (undefined) will be displayed for scale and offset if the two source channels are set to dissimilar units. Units are set using the channel’s Probe Units softkey. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 234
Measurements and Math Functions Example of √ (Square Root). Channel 1 √ waveform Square Source Math Math Displayed function root waveform waveform operator scale offset selected Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Aliasing happens when there are frequency components in the signal higher than half the sample rate. Since the FFT spectrum is limited by this frequency, any higher components are displayed at a lower (aliased) frequency. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 236
FFT. The FFT Menu provides four windows: Hanning, flattop, rectangular, and Blackman-Harris. For more information on leakage, see Keysight Application Note 243, “The Fundamentals of Signal Analysis” at http://literature.cdn.keysight.com/litweb/pdf/5952-8898E.pdf. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Window — selects a window to apply to your FFT input signal: • Hanning — window for making accurate frequency measurements or for resolving two frequencies that are close together. • Flat Top — window for making accurate amplitude measurements of frequency peaks. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 238
Pressing the FFT Preset softkey will automatically rescale the waveform and span and center will again automatically track the horizontal sweep speed setting. 3 To make cursors measurements, press the [Cursors] key and set the Source softkey to Math: f(t). Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 239
Once the FFT spectrum is displayed, the frequency span and center frequency controls are used much like the controls of a spectrum analyzer to examine the frequency of interest in greater detail. Place Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 240
For more information on the use of FFTs please refer to Keysight Application Note 243, ”The Fundamentals of Signal Analysis” at http://literature.cdn.keysight.com/litweb/pdf/5952-8898E.pdf. Additional information can be obtained from Chapter 4 of the book “Spectrum and Network Measurements” by Robert A. Witte. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
2 In the User Options Menu, select Precision to enable precision measurements and math waveforms. Precision measurements and math are disabled in the same way, or you can disable them by pressing [Save/Recall] > Default Setup. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 242
Measurements and Math Functions Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 243
Keysight InfiniiVision 6000L Series Oscilloscopes User's Guide 7 Printing and Saving Printing the Oscilloscope’s Display / 244 Quick Print / 244 To print the oscilloscope’s display / 245 Options / 245 Palette / 246 Saving Oscilloscope Data / 247 Selecting a Destination for Your Saved Data / 247...
When Quick Print is enabled, a Save function is performed immediately when the [Print] (or [Quick Print]) key is pressed. The save parameters must be set up in the Save/Recall Menu (see page 247) before pressing [Print]. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
• Invert Graticule Colors — The Invert Graticule Colors option can be used to reduce the amount of black ink it takes to print oscilloscope images by changing the black background to white. Invert Graticule Colors is the default mode. Graticule Not Inverted Graticule Inverted Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
• Color — When Color printing is selected, the traces are printed in color. • Grayscale — When Grayscale printing is selected, the traces are printed in shades of gray rather than in color. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
You can also save the oscilloscope’s display image using a web browser. See “Get NOTE Image" on page 287 for details. Selecting a Destination for Your Saved Data Press [Save/Recall]&Save to access the Save Menu and select a destination for saving your data. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 248
It can be used to recall oscilloscope setup and waveform trace files. The File Explorer in the Utility Menu can be used for loading files into the oscilloscope or deleting files. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Or, you can press the Spell softkey and press the Enter softkey to create the new file name. 3 You can press the Enter softkey or the Entry knob to move the cursor forward within the file name. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
• CSV data file — This creates a file of comma-separated variable values of all displayed channels and math waveforms. This format is suitable for spreadsheet analysis. Serial decode Lister data is also saved in CSV format. (See also page 414.) Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
The setup information will be sent to a separate file with a TXT extension. Invert Graticule Colors This setting is applicable when you have chosen BMP or PNG using the Format softkey. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 252
Only displayed data points are output (except when you are using segmented memory and the Save Segment control is set to All segments). Therefore, adjust the Horizontal controls to display the data you want to save. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 253
ASCII XY the data are presented in separate files for channel 1, channel 2, channel 3, channel 4, digital channels 0-7, digital channels 8-15, and math waveforms. For more information, see: • “Binary Data (.bin) Format" on page 406 Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 254
If you connect your serial bus signals to the oscilloscope as shown in the following table, the labels in the analysis software will be correct. If you do not, then you will have to re-map the signals in the analysis software. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
249 for details.) 4 Choose a file format using the Format softkey. For file format information see “Display Image and Waveform Data File Formats" on page 250. 5 Press the Press to Save softkey. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
4 Press the second softkey from the left and navigate to the oscilloscope’s internal memory location or the USB mass storage device that holds the saved trace and/or setup. See File Explorer (below) for more information. 5 Press the Press to Recall softkey. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
“USB0” and the second one is designated “USB5,” not “USB1.” This numbering method is normal; it is inherent in the USB driver. A file that has been deleted from the USB mass storage device cannot be recovered by the oscilloscope. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 258
Location when pointing to a directory that is currently selected, and Selected when pointing to a file that can be loaded or deleted. Pressing the Entry knob will initiate the selected activity. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 259
Printing and Saving Delete selected Return to Navigate Load selected file previous menu using Entry file knob, select using this key or push Entry knob Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 260
Printing and Saving Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 261
Keysight InfiniiVision 6000L Series Oscilloscopes User's Guide 8 Acquisition Modes To start and stop an acquisition / 262 To make a single acquisition / 264 Acquisition Modes / 265 Normal Mode / 265 Peak Detect Mode / 266 High Resolution Mode / 266...
Just press [Run/Stop] again. The acquisition will stop immediately and a partial waveform will be displayed. You can display the results of multiple acquisitions using infinite persistence. See “Infinite Persistence" on page 175. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 263
While running, maximizing the rate at which waveforms are drawn on the display provides the best representation of your input signal. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
(for example, if you are probing a DC level), set the trigger mode to Auto (see page 75) and press the [Single] key. If a trigger condition occurs, it will be used; if a trigger doesn’t happen, an untriggered acquisition will be taken. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
To select the acquisition mode press the [Acquire] key on the front panel. Normal Mode In Normal mode at slower sweep speeds, extra samples are decimated (in other words, some are thrown away). This mode yields the best display for most waveforms. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
High Resolution mode can be used on both single-shot and repetitive signals and it does not slow waveform update because the computation is done in the MegaZoom custom ASIC. High Resolution mode limits the oscilloscope's real-time bandwidth because it effectively acts like a low-pass filter. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
To use the Averaging mode 1 Press the [Acquire] key, then press the Acq Mode softkey until the Averaging mode is selected. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 268
The number of acquisitions being averaged is displayed in the # Avgs softkey. 1 Random noise on the displayed waveform 128 Averages used to reduce random noise See also “Trigger Noise Rejection" on page 79 Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Whenever both channels in a channel pair are on, the sample rate for all channels is halved. For example, when channels 1, 2, and 3 are on, the sample rate for all channels is 2 GSa/s. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 270
Realtime sampling on and 1 GHz when Realtime sampling is off. To see the sample rate, press the [Acquire] key on the front panel. The sample rate is displayed in the line just above the softkeys. Sample rate Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
The oscilloscope will continue to trigger until memory is filled, then the oscilloscope will stop. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 272
Segmented Memory mode is not selected. The Knob softkey also appears in the Horizontal Menu so you can make the same adjustments there. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 273
5 μs/div, and the Time Reference is set to Center, it will take at least 50 μs to fill all ten divisions and re-arm. (That’s 25 μs to capture pre-trigger data and 25 μs to capture post-trigger data.) Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 274
[Save/Recall]&Save&Settings&Save Seg. A USB storage device must be connected to the oscilloscope to enable this Save option. You can also save the screen image in BMP or PNG formats. For more information see “Save Segment" on page 253. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 275
Keysight InfiniiVision 6000L Series Oscilloscopes User's Guide 9 Web Interface Setting up the Oscilloscope’s LAN Connection / 277 Accessing the Web Interface / 281 Browser Web Control / 282 Get Image / 287 Identification Function / 288 Instrument Utilities / 289...
The Web browser must be Java-enabled with Sun Microsystems™ Java Plug-in. Before you can use the web interface, you must place the oscilloscope on the network and set up its LAN connection. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
IP, DNS IP, etc.) from your network administrator. 2 Connect the oscilloscope to the local area network (LAN) by inserting the LAN cable into the “LAN” port on the rear panel of the oscilloscope. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
These connection methods can be enabled or disabled later using the Configure Network tab in the web interface. 4 Open a web browser and enter the oscilloscope’s hostname in the address field. The oscilloscope’s Web page will be displayed. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
169.254.254.254; however, this could be different depending on negotiation during AutoIP. After the LAN connection is configured, you can find the oscilloscope’s IP address and use it to access the instrument’s Web interface: Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 280
9 When the instrument is found, select OK and OK to close the Add Instrument windows. Now, you can see the oscilloscope’s IP address in the Keysight Connection Expert and use it to access the instrument’s Web interface. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
279). It is possible to use a point-to-point connection, but using a normal LAN connection is the preferred method. 2 Type the oscilloscope’s hostname or IP address in the web browser. The oscilloscope’s web interface Welcome Page is displayed. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Remote Front Panel. After a few seconds the Remote Front Panel appears. 3 Use the Main Menu and the Function Keys to control the oscilloscope. To view Quick Help, right-click on a softkey. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 283
Web Interface Main Menu Function Keys Hint appears when you roll over with mouse Softkeys (Left-click to select, Right- click for Quick Help Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
281). 2 When the oscilloscope’s web interface is displayed, select Browser Web Control, then select Remote Programming. 3 On the Remote Programming page, click SCPI Commands..The SCPI Commands applet window appears. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 285
& b In the I/O Menu, press Control. c Use the Entry knob to point to LAN; then, press Control again to enable the LAN controller type. Multiple controller types may be enabled. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
For more information about connecting to the oscilloscope, refer to the Keysight Technologies USB/LAN/GPIB Interfaces Connectivity Guide. For a printable electronic copy of the Connectivity Guide, direct your Web browser to www.keysight.com and search for Connectivity Guide. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Welcome screen. After a delay of several seconds, the oscilloscope’s screen image will be displayed. 3 Right-click on the image and select “Save Picture As...” (or “Print Picture...”). 4 Select a storage location for the image file and click Save. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Identification off or press the OK softkey on the oscilloscope to continue. On 6000L Series oscilloscopes, the LAN indicator on the front of the instrument will flash green to identify the oscilloscope until you select Identification off. Identification Option Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
The Instrument Utilities page of the web interface lets you: • View installed options. • View firmware versions. • Install firmware upgrade files. • View calibration status. You can select these cababilities via a drop-down menu. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
“Accessing the Web Interface" page 281). 2 When the oscilloscope’s web interface is displayed, select the Configure Network tab from the instrument’s Welcome page. 3 Click the Modify Configuration button. Step 2 Step 1 Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 291
• On 6000L Series oscilloscopes, press the LAN RESET button on the front panel. • Using the web browser select the Configure Network tab, select Modify Configuration, erase the Password, and select Apply Changes. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 292
Web Interface Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Keysight InfiniiVision 6000L Series Oscilloscopes User's Guide 10 Serial Decode/Lister Serial Decode / 294 Lister / 295 CAN Serial Decode / 297 FlexRay Serial Decode / 305 C Serial Decode / 309 S Serial Decode / 313 LIN Serial Decode / 317...
[Utility] key, then press the Options softkey, the Licenses softkey, and the Show license information softkey. To order serial decode licenses, go to www.keysight.com and search for the product number (for example, N5432A) or contact your local Keysight Technologies representative (see www.keysight.com/find/contactus). Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Serial Decode" on page 297). 2 Press [Acquire]&Serial Decode&Lister. 3 Press Lister Display to enable the display. Before you can select a row or navigate through the Lister data, oscilloscope acquisitions must be stopped. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 296
Enable or disable the Track Time option. When enabled, as you select different Lister rows (using the Entry knob while acquisitions are stopped), the horizontal delay changes to the Time of the selected row. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Lister c In the Serial Decode Menu, press the Mode softkey. d Use the Entry knob (or repeatedly press and release the Mode softkey) to select the CAN serial decode mode. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 298
If you have connected one of the oscilloscope’s analog source channels to the CAN signal: Adjust the trigger level for the selected analog channel to the middle of the waveform by turning the Trigger Level knob. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 299
8 Set the trigger level. • For analog channels, turn the Trigger Level knob. • For digital channels, press the [D15-D0] key and the Thresholds softkey to access the threshold level setting softkeys. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 300
AutoTriggering. Press the [Mode/Coupling] key, then press the Mode softkey to set the trigger mode from Auto to Normal. You can use the Zoom horizontal sweep mode for easier navigation of the decoded data. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
• Angled waveforms show an active bus (inside a packet/frame). • Mid-level blue lines show an idle bus. • Decoded text is truncated at the end of the associated frame when there is insufficient space within frame boundaries. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
• The UTIL (bus load) indicator measures the percentage of time the bus is active. The calculation is done on 330 ms periods, approximately every 400 ms. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 303
ERR counter will be incremented. If a data frame contains an error that is not an active error it is considered a partial frame and no counters are incremented. Overload Error Frame Bus Utilization Total Frame Count Frame Count Count and Percentage and Percent Percent Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
• CRC — cyclic redundancy check. • Errors — highlighted in red. Errors can be Acknowledge (Ack, A), Form (Fo), or Frame (Fr). Different kinds of errors can be combined like “Fo,Fr” in the above example. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
2 If the decode line does not appear on the display, press the [Serial] key to turn it 3 If the oscilloscope is stopped, press the [Run/Stop] key to acquire and decode data. You can use the Zoom horizontal sweep mode for easier navigation of the decoded data. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
• Cycle number (decimal digits in yellow). • Data bytes (HEX digits in white). • Frame CRC (hex digits in blue plus red FCRC error message if invalid). • Frame/coding errors (specific error symbol in red). Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
The totalizer runs, counting frames and calculating percentages, even when the oscilloscope is stopped (not acquiring data). When an overflow condition occurs, the counter displays OVERFLOW. The counters can be reset to zero by pressing the Reset FlexRay Counters softkey. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
In the Serial Decode Menu, press the Mode softkey. d Use the Entry knob (or repeatedly press and release the Mode softkey) to select the I C serial decode mode. 2 Specify the I C signals: Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 310
If the setup does not produce a stable trigger, the I C signal may be slow enough that the NOTE oscilloscope is AutoTriggering. Press the [Mode/Coupling] key, then press the Mode softkey to set the trigger mode from Auto to Normal. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Read addresses appear in yellow along with the “R” character. • Restart addresses appear in green along with the “S” character. • Data values appear in white. • “A” indicates Ack (low), “~A” indicates No Ack (high). Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
• Restart — indicated with an “X”. • Address — colored blue for writes, yellow for reads. • Data — data bytes. • Missing Ack — highlighted in red, error if it has an “X”. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Use the Entry knob (or repeatedly press and release the Mode softkey) to select the I S serial decode mode. 2 Press the Settings softkey to open the I S Settings Menu. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 314
AutoTriggering. Press the [Mode/Coupling] key, then press the Mode softkey to set the trigger mode from Auto to Normal. You can use the Zoom horizontal sweep mode for easier navigation of the acquired data. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
• Red dots in the decode line indicate that more data can be displayed. Scroll or expand the horizontal scale to view the data. • Aliased bus values (undersampled or indeterminate) are drawn in red. • Unknown bus values (undefined or error conditions) are drawn in red. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
S Lister contains these columns: • Left Channel — displays the left channel data. • Right Channel — displays the right channel data. • Errors — highlighted in red and marked with an “X”. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
If you want to mask the upper two parity bits, ensure that the box under the Show Parity softkey is not selected. b If you want to include the parity bits in the identifier field, ensure that the box under the Show Parity softkey is selected. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 318
LIN bus signal. The LIN baud rate can be set to 2400 b/s, 9600 b/s, 10.4 kb/s, 19.2 kb/s, 115.2 kb/s, 625 kb/s, or User Defined. The default baud rate is 2400 b/s. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 319
LIN signal (this applies to MSO model oscilloscopes only): Press the [D15-D0] key and select Thresholds to set an appropriate threshold level for digital channels. The value of the trigger level or digital threshold is displayed in the upper-right corner of the display. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 320
AutoTriggering. Press the [Mode/Coupling] key, then press the Mode softkey to set the trigger mode from Auto to Normal. You can use the Zoom horizontal sweep mode for easier navigation of the decoded data. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
• Unknown bus values (undefined or error conditions) are drawn in red. • If there is an error in the synch field, SYNC will appear in red. • If the header exceeds the length specified in the standard, THM will appear red. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
The base setting is used for the display of the remote terminal address and the data, in both the decode line and in the Lister. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
• Command/Status or Data words with a Parity error have the decode text displayed in red instead of green or white. • SYNC errors are displayed with the word "SYNC" within red angle brackets. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
• RTA – displays the Remote Terminal Address for Command/Status words, nothing for Data words. • Word Type – “Cmd/Status” for Command/Status words, “Data” for Data words. For Command/Status words the background color is green to match the decode text color. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 326
Data word. • Errors – “Sync”, “Parity”, or “Manchester” errors as appropriate. The background color is red to indicate an error. For more information on using the Lister display, see “Lister" on page 295. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Lister c In the Serial Decode Menu, press the Mode softkey. d Use the Entry knob (or repeatedly press and release the Mode softkey) to select the SPI serial decode mode. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 328
Changing SPI signals in the Serial Decode setup also changes them in the Trigger setup. a Press the Settings softkey to access the SPI Settings Menu. b Press the Signals softkey to access the SPI Signals Menu. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 329
If you have connected one of the oscilloscope’s analog source channels to the Clock signal: Adjust the trigger level for the selected analog channel to the middle of the waveform by turning the Trigger Level knob. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 330
Entry knob to select the number of bits in a word. 4 If the decode line does not appear on the display, press the Decode softkey to turn it on. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
• Angled waveforms show an active bus (inside a packet/frame). • Mid-level blue lines show an idle bus. • The number of clocks in a frame appears in light-blue above the frame, to the right. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
• Unknown bus values (undefined or error conditions) are drawn in red. Interpreting SPI Lister Data In addition to the standard Time column, the SPI Lister contains these columns: • Data — data bytes (MISO or MOSI). Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Use the Entry knob (or repeatedly press and release the Mode softkey) to select the UART/RS232 serial decode mode if it is not already selected. Changing UART/RS232 signals in the Serial Decode setup also changes them in the Trigger NOTE setup. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 334
Press the Signals softkey, and select the oscilloscope channels that you have connected to the Rx and Tx signals. Set the trigger levels. See “UART/RS232 Trigger" on page 156 for detailed instructions. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 335
However, if a parity error occurs the data will be displayed in red. 8 Make sure the trigger or threshold levels are set to the middle of the signals: • For analog channels, turn the Trigger Level knob. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 336
AutoTriggering. Press the [Mode/Coupling] key, then press the Mode softkey to set the trigger mode from Auto to Normal. You can use the Zoom horizontal sweep mode for easier navigation of the acquired data. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Expand the horizontal scale to allow the data to display. • An unknown (undefined) bus is shown in red. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
When an overflow condition occurs, the counter displays OVERFLOW. The counters can be reset to zero by pressing the Reset UART Counters softkey. Rx Frame Error Frame Tx Frame Count Count Count and Percent Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
In addition to the standard Time column, the UART/RS232 Lister contains these columns: • Rx — receive data. • Tx — transmit data. • Errors — highlighted in red, Parity Error or Unknown Error. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 340
Serial Decode/Lister Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 341
Keysight InfiniiVision 6000L Series Oscilloscopes User's Guide 11 Mask Test To create a mask from a “golden” waveform (Automask) / 342 Setup Options / 344 Mask Statistics / 347 To manually modify a mask file / 349 Building a Mask File / 353 One way to verify a waveform's compliance to a particular set of parameters is to use mask testing.
6 Adjust the mask’s horizontal tolerance (± Y) and vertical tolerance (± X). These are adjustable in graticule divisions or in absolute units (volts or seconds), selectable using the Units softkey. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
If you press Create Mask and it appears that no mask was created, check the ± Y and ± X settings. They may be set so large that the mask is not visible. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
The oscilloscope will stop when the first error is detected (on the first waveform that does not conform to the mask). This setting supersedes the Minimum # of Tests and Minimum Time settings. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Automask, then press Create Mask. The Source softkey in the Mask Setup Menu is the same as the Source softkey in the Automask Menu. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
The Pass/fail status is evaluated periodically. When the evaluation of the testing period results in a pulse output, the trigger output pulses high (+ 5 V). Otherwise, the trigger output remains at low (0 V). Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
• Accumulated test time (in hours, minutes, seconds, and tenths of seconds). And for each analog channel: • Number of failures (acquisitions in which the signal excursion went beyond the mask). • Failure rate (percentage of failures). Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Enable Transparent mode to write measurement values and statistics on the screen without a background. Disable the Transparent mode to show them with a light blue background. This controls the display of statistics for both Quick Meas and the optional Mask test feature. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Example: autoMask CH1 OCT 03 09:40:26 2008 When a mask file contains the keyword “autoMask” in the title, the edge of the mask is passing by definition. Otherwise, the edge of the mask is defined as a failure. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 350
Similarly, the vertices in Region 2 describe the line that forms the top of the bottom part of the mask. The vertices in a mask file are normalized. There are four parameters that define how values are normalized: • X1 Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 351
The mask scaling controls how the normalized vectors are interpreted. This in turn controls how the mask is drawn on the display. The remote programming commands that control mask scaling are: :MTES:SCAL:BIND 0 :MTES:SCAL:X1 -400.000E-06 :MTES:SCAL:XDEL +800.000E-06 Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 352
Mask Test :MTES:SCAL:Y1 +359.000E-03 :MTES:SCAL:Y2 +2.35900E+00 Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
/* Region Number */ 1 /* Number of vertices */ 4 -12.50, -10.00, 1.750 10.00, 1.750 12.50, /* Region Number */ 2 /* Number of vertices */ 5 -10.00, 1.000 -12.50, 0.500 -15.00, 0.500 -15.00, 1.500 Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 354
/* Region Number */ 8 /* Number of vertices */ 4 -12.50, -10.00, -1.750 10.00, -1.750 12.50, setup :MTES:ENAB 1 :CHAN1:RANG +4.00E+00;OFFS +0.0E+00;COUP DC;IMP ONEM;DISP 1;BWL 0;INV 0 :CHAN1:LAB "1";UNIT VOLT;PROB +1.0E+00;PROB:SKEW +0.0E+00;STYP SING Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 355
The number of acquisitions tested is also logged and displayed as “# of Tests”. The mask file allows greater resolution than the 256 X 1000 database. Some quantization of data occurs to reduce the mask file data for display on-screen. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 356
Mask Test Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Keysight InfiniiVision 6000L Series Oscilloscopes User's Guide 12 Controls and Connectors 6000L Series Front and Rear Panel / 358 Front Panel Control and Connector Descriptions / 359 Rear Panel Control and Connector Descriptions / 361...
12. GPIB 3. 10 MHz 6. XGA 5. Digital 7. USB 8. USB External Trigger Port Connector Calibration Reference Video Input Host Port Device Trigger Output Protect Connector Output Connector Port Input Switch Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
4. Channel Input BNC Connector Attach the oscilloscope probe or BNC cable to the BNC connector. This is the channel’s input connector. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 360
The LAN status indicator flashes green when the identify function is activated (see “Identification Function" on page 288). 27. Reset Switch (6000L models only.) When power is on, press and release this recessed pushbutton to default LAN parameters. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
This port is for controlling the oscilloscope via GPIB. You can issue remote commands from a host computer and retrieve data from the oscilloscope. See “Remote Programming with Keysight IO Libraries" on page 286. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 362
Controls and Connectors Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 363
Keysight InfiniiVision 6000L Series Oscilloscopes User's Guide 13 Digital Channels To connect the digital probes to the device under test / 364 Acquiring waveforms using the digital channels / 368 To display digital channels using AutoScale / 369 Interpreting the digital waveform display / 371...
Probe cable for digital channels CAUTION Use only the Keysight logic probe and accessory kit supplied with the mixed-signal oscilloscope. 3 Route the cable under the oscilloscope and bring it out in front. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 365
The ground lead improves signal fidelity to the oscilloscope, ensuring accurate measurements. Channel Pod Ground Circuit Ground 5 Connect a grabber to one of the probe leads. (Other probe leads are omitted from the figure for clarity.) Grabber Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 366
7 For high-speed signals, connect a ground lead to the probe lead, connect a grabber to the ground lead, and attach the grabber to ground in the device under test. Signal Lead Ground Lead Grabber Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 367
Digital Channels 8 Repeat steps 3 through 6 until you have connected all points of interest. Signals Ground Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
For digital channels, each time the oscilloscope takes a sample it compares the input voltage to the logic threshold. If the voltage is above the threshold, the oscilloscope stores a 1 in sample memory; otherwise, it stores a 0. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Install probe clips on channels 0 and 1 of the digital probe cable. Connect the probes for digital channels 0 and 1 to the Probe Comp output on the front panel of the instrument. Be sure to connect the ground lead to the ground lug beside the Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 370
[Save/Recall] key, then the Default Setup softkey. Then press the [AutoScale] key. You should see a display similar to the following. Figure 2 AutoScale of digital channels 0 through 7 (MSO models only) Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
A digital channel can be always high ( ), always low ( ), or actively toggling logic states ( ). Any channel that is turned off will be grayed out in the activity indicator. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
2 Press the size ( ) softkey to select how the digital channels are displayed. The sizing control lets you spread out or compress the digital traces vertically on the display for more convenient viewing. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
1 With the Digital Channel Menu displayed, rotate the Entry knob to select the desired channel from the popup menu. 2 Press the softkey that is directly below the popup menu to switch the selected channel on or off. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
[D15-D0] key twice to switch the digital channels off. The first push displays the Digital Channel Menu, and the second push switches the channels off. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
2 Press the Turn off (or Turn on) softkey for the D15 - D8 group or the D7 - D0 group. Each time you press the softkey, the softkey’s mode toggles between Turn on and Turn off. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
If the Thresholds softkey is set to User, press the User softkey for the channel group, then turn the Entry knob to set the logic threshold. There is one User softkey for each group of channels. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
D designation (where nn is a one or two digit channel number from 0 to 15) to D * . The “ * ” indicates that two channels are overlaid. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
(see page 379). The Base softkey lets you choose to display the bus values in hex or binary. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 379
1 Turn on Cursors (by pressing the [Cursors] key on the front panel) 2 Press the cursor Mode softkey and change the mode to Hex or Binary. 3 Press the Source softkey and select Bus1 or Bus2. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 380
The dollar sign ($) will be displayed in the bus value when the bus value cannot be displayed as a hex value. This occurs when one or more “don’t cares” (X) are combined with low (0) and high (1) logic levels in the pattern specification, or Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 381
A byte that consists of all don’t cares (X) will be displayed in the bus as a don’t care (X). Digital channel values at cursor Analog channel values at cursor Trigger pattern definition Bus values displayed page 126 for more information on Pattern triggering. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
(see the following figure). This high-frequency model takes into account pure tip capacitance to ground as well as series tip resistance, and the cable’s characteristic impedance (Z DC and Low-Frequency Probe Equivalent Circuit 8.5 pF 100 kW Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 383
They are designed to provide as much series tip resistance as possible. Stray tip capacitance to ground is minimized by the proper mechanical design of the probe tip assembly. This provides the maximum input impedance at high frequencies. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
(dt), the common mode voltage may increase to a level that causes false data generation. Common Mode Input Voltage Model Probe 1 L (GND) Probe (Common Mode) Ground Probe 2 Probe N Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
This probe consists of a 2 m logic analyzer probe cable and a 01650-63203 termination adapter that provides the proper RC networks in a very convenient package. Three 20-pin, low-profile, straight board connectors are included. Additional board connectors can be ordered from Keysight Technologies. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Keysight InfiniiVision 6000L Series Oscilloscopes User's Guide 14 Utilities To restore the oscilloscope to its default configuration / 388 To perform service functions / 389 User Calibration / 390 Self Test / 393 This chapter explains oscilloscope utilities.
Acquire mode normal, [Run/Stop] to Run, cursors and measurements off. Labels All custom labels that you have created in the Label Library are preserved (not erased), but all channel labels will be set to their original names. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
The Service Menu lets you: • Perform User Calibration on the oscilloscope. • View User Calibration status. • Perform instrument Self Test. • View information about your oscilloscope model number, code revision information, and User Calibration status. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
2 Connect short (12 inch maximum) equal length cables to each analog channel’s BNC connector on the front of the oscilloscope. You will need four equal-length cables for a 4-channel oscilloscope. Use 50Ω RG58AU or equivalent BNC cables when performing User Cal. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 391
4 Press the [Utility] key, then press the Service softkey. 5 Begin the Self Cal by pressing the Start User Cal softkey. 6 When the User Cal is completed, set the rear-panel CALIBRATION switch to PROTECTED. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 392
Note that passive probes do not need to be calibrated, but InfiniiMax probes can be calibrated. For more information about calibrating probes see page Results: User Cal date: Change in temperature since last User Cal: Failure: Comments: Probe Cal Status: Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
• to verify proper operation after the oscilloscope has been repaired. Successfully passing Self Test does not guarantee 100% of the oscilloscope's functionality. Self Test is designed to provide an 80% confidence level that the oscilloscope is operating properly. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 394
Utilities Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Reference About Oscilloscope Press [Utility]&Service&About Oscilloscope to display information about your oscilloscope model number, serial number, software version, boot version, graphics version, and installed licenses. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Reference Software and Firmware Updates From time to time Keysight Technologies releases software and firmware updates for its products. To search for firmware updates for your oscilloscope, direct your web browser to www.keysight.com/find/6000sw. To view the currently installed software and firmware press [Utility]&Service&About Oscilloscope.
• SGM — Segmented memory. This license lets you capture infrequent or burst signals with great resolution by eliminating the capture of your signal’s “dead time.” Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 399
Reference • SND — I S serial decode option (for 4 channel or 4+16 channel models only). Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
16 time-correlated digital timing channels. To view the currently installed licenses press [Utility]&Options&Licenses&Show license information. For information about upgrading your oscilloscope through licensing, contact your local Keysight Technologies representative or see www.keysight.com/find/6000. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
When installed, the secure environment mode license will appear as SEC in the Installed Licenses line of the “About Oscilloscope” display. To access the About Oscilloscope display, press the [Utility] key, then press the Service softkey, then press the About Oscilloscope softkey. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Do not apply more than ±15 V at the 10 MHz REF BNC connector on the rear panel or damage to the instrument may occur. 2 Press [Utility]&Options&Rear Panel&Ref Signal. 3 Use the Entry Knob and the Ref Signal softkey to select 10 MHz input. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 403
The oscilloscope will resume sampling when the externally supplied sample clock becomes stable again. 10 MHz unlock icon and system stopped Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
3 Press [Utility]&Options&Rear Panel&Ref Signal. 4 Use the Entry Knob and the Ref Signal softkey to select 10 MHz output. The oscilloscope will output it’s 10 MHz reference signal at TTL levels. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
2 Clean the external surfaces of the oscilloscope with a soft cloth dampened with a mixture of mild detergent and water. 3 Make sure that the instrument is completely dry before reconnecting it to a power source. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
(including math and digital channels, with up to seven waveform for each digital pod). When saving all segments the number of waveforms is equal to the number of active channels multiplied by the number of segments acquired. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
A 32-bit integer that is the number of bytes in the header. Waveform Type A 32-bit integer that is the type of waveform stored in the file: • 0 = Unknown. • 1 = Normal. • 2 = Peak Detect. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 408
64-bit floating point number. If the value is zero then no data has been acquired. X Units A 32-bit integer that identifies the unit of measure for X values in the acquired data: • 0 = Unknown. • 1 = Volts. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 409
Buffer Type A 16-bit short that is the type of waveform data stored in the file: • 0 = Unknown data. • 1 = Normal 32-bit float data. • 2 = Maximum float data. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
A 32-bit integer that is the size of the buffer required to hold the data points. Example Program for Reading Binary Data To find an example program for reading binary data, direct your web browser to www.keysight.com/find/6000sw, and select “Example Program for Reading Binary Data”. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Waveform Header N Number of Waveform Buffers = 1 140 bytes Waveform Data Buffer Type = 1 (floating point) Header N Bytes per Point = 4 12 bytes Voltage Data N buffer size Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 412
Waveform Header 2 Number of Waveform Buffers = 1 140 bytes Waveform Data Buffer Type = 6 (unsigned char) Header 2 Bytes per Point = 1 12 bytes Pod 2 Timing Data buffer size Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Page 413
140 bytes Time Tag = time between segment 1 and N Waveform Data Buffer Type = 1 (floating point) Header N Bytes per Point = 4 12 bytes Voltage Data N buffer size Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
CSV file will be the best-estimate value at each of the 1000 horizontal points. Therefore, the min and max values may not appear in the CSV file. This occurs when oversampling occurs (10 * seconds per division * maximum sample rate > 1000). Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
The license is located on the Keysight InfiniiVision Oscilloscopes Documentation CD-ROM. RealVNC source code can be obtained from RealVNC or by contacting Keysight. Keysight will charge for the cost of physically performing the source distribution. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
2 Enter your product’s model number and serial number. The system will search for the warranty status of your product and display the results. If the system cannot find your product’s warranty status, select Contact Us and speak with an Keysight Technologies representative. Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
• Description of service required or failure indication 2 Remove accessories from the oscilloscope. Only return accessories to Keysight Technologies if they are associated with the failure symptoms. 3 Package the oscilloscope. You can use the original shipping container, or provide your own materials sufficient to protect the instrument during shipping.
Page 424
2, TTL threshold, waveform TV trigger, cursor tracking, intensity, printing, recall, UART totalizer, reference point, UART trigger, save options, UART/RS232 frame counter, save to scope memory, UART/RS232 license, save to USB, Keysight InfiniiVision 6000L Series Oscilloscopes User’s Guide...
Need help?
Do you have a question about the InfiniiVision 6000L Series and is the answer not in the manual?
Questions and answers