Power:signals:vmaximum:onoff:off - Keysight Technologies InfiniiVision 4000 X Series Programmer's Manual

Hide thumbs Also See for InfiniiVision 4000 X Series:
Table of Contents

Advertisement

28
:POWer Commands

:POWer:SIGNals:VMAXimum:ONOFf:OFF

Command Syntax
:POWer:SIGNals:VMAXimum:ONOFf:OFF <value>[suffix]
<value> ::= Maximum expected input Voltage in NR3 format
[suffix] ::= {V | mV}
The :POWer:SIGNals:VMAXimum:ONOFf:OFF command specifies the maximum
expected input voltage. This value is used to set the vertical scale of the channel
probing voltage for turn off analysis.
When the :POWer:ITYPe is DC, this command defines the maximum DC input
voltage amplitude value. The values can be negative.
When the :POWer:ITYPe is AC, this command defines the maximum peak-to-peak
input voltage. Only positive values are allowed.
Query Syntax
:POWer:SIGNals:VMAXimum:ONOFf:OFF?
The :POWer:SIGNals:VMAXimum:ONOFf:OFF query returns the expected
maximum input voltage setting.
Return Format
<value><NL>
<value> ::= Maximum expected input Voltage in NR3 format
See Also
788
(see
page
1610)
":POWer:ITYPe"
on page 750
":POWer:ONOFf:APPLy"
":POWer:SIGNals:AUTosetup"
":POWer:SIGNals:DURation:ONOFf:OFF"
":POWer:SIGNals:IEXPected"
":POWer:SIGNals:OVERshoot"
":POWer:SIGNals:VSTeady:ONOFf:OFF"
":POWer:SIGNals:SOURce:CURRent<i>"
":POWer:SIGNals:SOURce:VOLTage<i>"
on page 754
on page 776
on page 781
on page 785
on page 786
on page 790
on page 793
on page 794
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the InfiniiVision 4000 X Series and is the answer not in the manual?

This manual is also suitable for:

Infiniivision mso-x 4034aInfiniivision mso-x 4054aInfiniivision mso-x 4032aInfiniivision mso-x 4022aInfiniivision mso-x 4104aInfiniivision mso-x 4154a ... Show all

Table of Contents