:POWer:CLResponse:DATA:GMARgin:FREQuency
Query Syntax
:POWer:CLResponse:DATA:GMARgin:FREQuency?
After the Control Loop Response (Bode) power analysis has been performed (see
:POWer:CLResponse:APPLy), the
:POWer:CLResponse:DATA:GMARgin:FREQuency? query returns the 0° phase
crossover frequency in Hz.
Return Format
<frequency><NL>
<frequency> ::= 0 degrees phase crossover frequency in Hz in NR3 format
The query returns +9.9E+37 if the value cannot be calculated from the last sweep
(that is, if there are no crossover points).
See Also
•
•
•
•
•
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
(see
page
1610)
":POWer:CLResponse:APPLy"
":POWer:CLResponse:DATA"
":POWer:CLResponse:DATA:GMARgin"
":POWer:CLResponse:DATA:PMARgin"
":POWer:CLResponse:DATA:PMARgin:FREQuency"
on page 714
on page 715
on page 716
on page 718
:POWer Commands
on page 719
28
717
Need help?
Do you have a question about the InfiniiVision 4000 X Series and is the answer not in the manual?