:SBUS<n> Commands
:SBUS<n>:LIN:STANdard
Command Syntax
:SBUS<n>:LIN:STANdard <std>
<std> ::= {LIN13 | LIN13NLC | LIN20}
The :SBUS<n>:LIN:STANdard command sets the LIN standard in effect for
triggering and decoding:
• LIN13 — LIN 1.3.
• LIN13NLC — LIN 1.3 (no length control). Select this for systems where length
• LIN20 — LIN 2.X.
For LIN 1.2 signals, use the LIN 1.3 setting. The LIN 1.3 setting assumes the signal
follows the "Table of Valid ID Values" as shown in section A.2 of the LIN
Specification dated December 12, 2002. If your signal does not comply with the
table, use the LIN 2.X setting.
Query Syntax
:SBUS<n>:LIN:STANdard?
The :SBUS<n>:LIN:STANdard? query returns the current LIN standard setting.
Return Format
<std><NL>
<std> ::= {LIN13 | LIN13NLC | LIN20}
See Also
980
(see
page
1610)
control is not used and all nodes have knowledge of the data packet size. In LIN
1.3, the ID may or may not be used to indicate the number of bytes. (In LIN 2.X,
there is no length control.)
"Introduction to :TRIGger Commands"
":TRIGger:MODE"
on page 1308
":SBUS<n>:LIN:SIGNal:DEFinition"
":SBUS<n>:LIN:SOURce"
on page 1295
on page 1543
on page 979
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the InfiniiVision 4000 X Series and is the answer not in the manual?

Subscribe to Our Youtube Channel

This manual is also suitable for:

Infiniivision mso-x 4034aInfiniivision mso-x 4054aInfiniivision mso-x 4032aInfiniivision mso-x 4022aInfiniivision mso-x 4104aInfiniivision mso-x 4154a ... Show all

Table of Contents