82 . DSC 4000 Installation and Hardware Guide
Error message ID
ERR_FURN_NOT_EMPTY
ERR_NOT_AT_SENS
ERR_NO_ZERO_DETECT
Error
Furnace still contains a
sample while trying to
load a new one. This
error can be generated
only after loading at
least one sample after
power on, as the
instrument cannot detect
if a sample is present in
the furnace from a
previous power on
period
Sensor position not
reached during
autocalibrating the
sensor position
Zero position is not
detected
Possible
Corrective
Cause
Action
Failure picking
Check if gripper
sample out of
fingers are bent.
furnace
Replace gripper
assembly
Incorrect
horizontal
Adjust baseplate
baseplate
adjustment
Incorrect
Recalibrate the
horizontal
AS 6000
calibration
AS 6000 run
Reset parent
was interrupted
instrument and
after putting a
setup link again
sample in
furnace and
restarting
Vertical stepper
Check the DSC
missed steps
lower lid position
because it was
blocked while
running at slow
speed. (During
slow gripper
speed, the
stepper blocked
detection is
disabled.)
No detection of
Do not place
the gripper
instrument in
fingers because
direct sunlight
of direct
sunlight at the
light switches
Gripper
Align gripper
assembly
assembly
displaced