12. Enter the name of the reference material used, the Expected Onset value, the
Measured Onset value just measured for each reference material, and click the check
box in the Use column.
13. Select the Heat Flow tab.
14. Enter the name of the reference material used (we recommend indium for this
calibration), the Expected
Weight of the reference material used.
15. Click the Save and Apply button. In the Save As dialog box, select the default.d6c
calibration file and click OK. Answer Yes to overwrite the existing file to save the new
calibration values.
16. Click Close to exit Calibration and send the new calibration values to the DSC 4000.
The DSC 4000 calibration is complete.
This completes the installation of the DSC 4000.
H value, the Measured
Install a DSC 4000 . 51
H result just measured, and the