30 . DSC 4000 Installation and Hardware Guide
Care must be taken when handling the outer furnace lid so its ceramic
CAUTION
lining is not damaged.
ATTENTION
Des précautions doivent être prises lors de la manipulation du couvercle
du four extérieur, de sorte que sa doublure en céramique n'est pas
endommagée..
Avoid contamination of the furnace, use high-purity purge gases to purge any reaction
products out of the furnace, and heat the furnace to temperatures above 430 °C ONLY
when absolutely necessary.
Check that the cooling liquid supply is circulating before starting a run.
Ensure the purge gases are set to the recommended flow rates. A rate that is too high
will disturb the inner furnace lid.
Do not operate the instrument in a cold room. The ambient temperature and the
temperature of the instrument should be between 10 °C and 30 °C.
Never open the furnace when it is at an elevated temperature.
Important Specifications
Maximum power consumption
Supply voltage
Supply voltage frequency
Fuses
Safe temperature range
Dimensions L x H x D
Weight
430 VA
100–120 V or 220–240 V
50–60 Hz
6.3 AT or 3.15 AT
10–40 °C
38 x 17 x 41 cm (H 29 cm with autosampler)
17 kg (19 kg with autosampler)