Error Messages From The Dsc 4000 - PerkinElmer DSC 4000 Installation And Hardware Manual

Thermal analysis
Table of Contents

Advertisement

Error Messages from the DSC 4000

Error message ID
ERR_DUE_TO_AUTOSAMPLER
ERR_AUTOSAMPLER_RESPONS
ERR_AUTOSAMPLER_I2C_WD
Error
Possible Cause
AS 6000 entered
Refer to AS 6000
fatal error state.
error message
This caused DSC
to enter fatal
error state
Timeout on
I2C connection
command.
cable defective
Response from
command
transmitted to
AS 6000
No link between
I2C connection
DSC and AS 6000
cable defective
via I2C
AS 6000 Autosampler . 83
Corrective
Action
Refer to AS 6000
troubleshooting
(above)
Replace I2C cable
Replace I2C cable