76 . DSC 4000 Installation and Hardware Guide

Running a Play List with the AS 6000

Pyris Player is the backbone of Pyris Software for Windows automation. It was created with
autosamplers in mind. In addition to the standard play list items — Load Sample, Run
Method, Return Sample — there is also a Sample Group. The Sample Group simplifies
grouping like samples together (as you would have in a sample tray of the autosampler).
These like samples use the same test method and data analysis as part of the Sample Group.
A Sample Group consists of a Sample List and a Data Analysis List. A specific method is
selected for the samples in the Sample List. The Data Analysis List allows you to access all
necessary functions for data recall, curve manipulation, optimization, and calculations for
automatic data analysis.
Before starting a run, perform the steps below:
NOTE: Pyris play lists created using previous version of Pyris software are usually compatible
in later versions of Pyris. And play lists created for one instrument may be compatible
with other instruments of the same type. However, play lists created in later versions
of software may not be compatible with earlier versions of software.
1. Review the safety and warning notes for the analyzer.
2. Verify that the purge gas (if you are using it) and cooling device tubes are properly
connected at the rear of the instrument.
3. Turn on the gas and adjust the pressure.
4. Turn on the chiller and adjust according to its instructions.
5. Verify that the electrical and cable connections between the computer and the
instrument and other components of the system are properly connected.
6. Turn on the power to the system components.
7. Prepare your samples.
8. Start Pyris software and click the parent instrument's button in the Pyris Manager. Click
the Pyris Player button on the toolbar:
new one.
There are many ways to use the play list with the autosampler. A quick way to create a
play list for an STA 6000 or a TGA 4000 using the Sample Group feature is given below:
Load empty sample pans into the locations you want to use in the sample tray.
Create a new play list that contains only a Sample Group. Click the Tare All
button. In the Advanced Tare Options dialog box, select This Group Only;
Populate from Tray.
. Either open an existing play list or create a