46 . DSC 4000 Installation and Hardware Guide
Connecting to the Intracooler for the DSC 4000:
Frosting is possible on the rear of the DSC 4000. To prevent this problem
CAUTION
the exhaust tube must be properly installed on the analyzer.
ATTENTION
Le givrage est possible à l'arrière du DSC 4000. Pour éviter ce problème,
le tube d'échappement doit être correctement installé sur l'analyseur.
Exhaust tube installation:
•
For the KTS Kinetics Cooler use the entire piece as it does fit over the hose.
•
For the PolyScience Intracooler use a piece that slips over the two chiller pipes,
approximately 3.5 cm long.
Precautions for Intracoolers:
Do not lift or move the unit by pulling the flexible line.
•
Do not make any sharp bends in the flexible line.
•
•
Do not block or restrict air flow to the vents on the unit.
•
Do not operate the unit in temperatures above 32 °C.
•
Do not operate at a line voltage outside of 110 V +/-10% or 220 V +/-10% when
operating at 50/60 Hz.