Set Up the DSC 4000 System Components

Place the DSC 4000 on your laboratory bench. Do not plug in the power cord until instructed.
The computer and printer should already be on the bench and installed according to each
item's instructions.
Be sure to set up all of the system components listed below before switching on the
DSC 4000:
Select the correct voltage
Install the purge gases
Install the cooling device

Select the Correct Voltage for the DSC 4000

The DSC 4000 accepts the following voltages: 100–120 V and 200–240 V (50/60 Hz).
It is configured and shipped to operate at 200–240 V. Selection of the correct voltage for
your lab can be done at the AC inlet plug at the rear of the DSC 4000.
First check whether the voltage selection corresponds to the voltage in your laboratory. This
can be seen easily: the voltage range with the text in an upright position and the arrow
pointing at the small line represents the voltage that has been selected.
To select the correct voltage for the DSC 4000 in your laboratory:
WARNING
AVERTISSEMENT
Ensure that the m ains pow er lead is disconnected from the
instrum ent before proceeding.
Assurez-vous que le cordon d'alim entation secteur est
débranché de l'instrum ent avant de continuer.
Install a DSC 4000 . 39