Connect The Dsc 4000 System Components - PerkinElmer DSC 4000 Installation And Hardware Manual

Thermal analysis
Table of Contents

Advertisement

Connect the DSC 4000 System Components

The computer and optional printer should already be connected together and the Pyris
software already installed. To complete the connection of the system components, the
DSC 4000 has to be connected to the computer.
Make sure that the computer is off before plugging in the DSC 4000.
CAUTION
Assurez-vous que l'ordinateur est éteint avant de brancher le DSC 4000.
ATTENTION
The image below shows the connections on the DSC 4000:
Make sure all components are turned off and unplugged before you begin.
1. Connect one end of the RS-232 cable to the back of the DSC 4000.
2. Connect the other end of the RS-232 cable to COM1 on the back of the computer.
3. Connect the printer and printer cable to the computer.
4. Connect the analyzer, printer and computer power cords to AC outlets.
5. Once the system connections are made, switch on the system in the following order:
Computer
DSC 4000
Printer
6. Turn on the purge gas and dry gas, the water for the DSC 4000 cooling system or any
other accessories for the cooling device you are using.
Install a DSC 4000 . 47

Advertisement

Table of Contents
loading

Table of Contents