Welch Allyn Acuity Directions For Use Manual page 132

Central monitoring systems
Hide thumbs Also See for Acuity:
Table of Contents

Advertisement

128
Chapter 7 Move a monitored patient
To switch a hardwired Propaq CS monitor to a wireless connection
Throughout this procedure, leave the monitor turned on.
1. In the upper left corner of the Propaq CS monitor display,
touch NET OFF.
2. Within 15 seconds, remove the Acuity System network cable
from the Propaq monitor.
The CONFIRM PATIENT? prompt appears when the monitor connects with an Acuity
System.
3. Take one of these steps:
WARNING When a hardwired Propaq CS monitor is disconnected from an
Acuity System, if the monitor has wireless capability, it attempts to connect
wirelessly to an Acuity System.
After connection, go to the Acuity Central Station where you want the monitor to
have connected. Ensure that the patient monitor's icon appears on the central
station map.
If it does not, contact your Biomedical Engineering department. If you are a
biomedical engineer, see
on page 165.
If the information in the CONFIRM PATIENT? prompt is correct, click YES.
At the central station, the patient remains confirmed.
If the patient is moving, see
CS monitor"
on page 130.
If the information in the CONFIRM PATIENT? prompt is incorrect, click NO and
see
"To respond to the CONFIRM PATIENT? prompt"
Welch Allyn Acuity Central Monitoring System
"Identify or change a monitor's network name (ESSID)"
"To arrange a patient move from a wireless Propaq
on page 29.