Welch Allyn Acuity Directions For Use Manual page 139

Central monitoring systems
Hide thumbs Also See for Acuity:
Table of Contents

Advertisement

Directions for use
Permanently stop monitoring
To permanently stop monitoring a patient
1. From the monitor's Main Menu, take these
steps.
For Propaq CS and Encore monitors:
a. In the top left or right corner of the monitor, press NET OFF.
b. Turn off the monitor.
For a Micropaq monitor:
a. Press
b. Select END TELE.
c. Remove the battery from the Micropaq monitor.
For a Propaq LT monitor:
a. Press the arrow keys to underline
b. Select Acuity menu.
c. Select END TELE.
d. Turn off the monitor.
e. Select Delete & Shut Down.
For a Welch Allyn 1500 Patient Monitor:
a. Press the Standby key and select Discontinue Monitoring.
b. Turn off the monitor.
c. At Acuity, select END TELE.
Note
2. Discharge the patient from Acuity System monitoring.
The monitor disconnects from the Acuity System.
The trend information for this patient is erased from the monitor.
to highlight ACUITY, and then press
The monitor disconnects from the Acuity System.
screen, and then press
The monitor disconnects from the Acuity System.
The trend information for this patient is erased from the monitor.
The patient is discharged from the Acuity System and the
monitor enters standby mode.
or
Turning off the monitor is optional and not required when selecting END TELE.
In either case, the trend information for this patient is erased from the monitor.