Welch Allyn Acuity Directions For Use Manual page 136

Central monitoring systems
Hide thumbs Also See for Acuity:
Table of Contents

Advertisement

132
Chapter 7 Move a monitored patient
5. Enter or confirm the patient information.
For instructions, see
on page 40.
To move a patient from a Propaq LT monitor to a different monitor
1. On the Propaq LT monitor, press the arrow keys to underline
the top of the screen, and then press
2. Select Acuity menu.
3. Select END TELE.
The monitor disconnects from the Acuity System.
4. Turn off the monitor.
5. Select Delete & Shut Down.
The trend information for this patient is erased from the monitor.
6. Connect the patient to a new monitor, and connect the monitor to the Acuity System.
At the central station, the patient's yellow icon appears in an unconfirmed state on
the map.
7. Enter or confirm the patient information.
For instructions, see
on page 40.
To move a patient from a wireless Propaq CS monitor to a different monitor
1. On the Propaq CS monitor, touch NET OFF.
2. Turn off the monitor.
The trend information for this patient is erased from the monitor.
3. Connect the patient to a new monitor and connect the monitor to the Acuity System.
At the central station, the patient's yellow icon appears in an unconfirmed state on
the map.
4. Enter or confirm the patient information.
For instructions, see
on page 40.
"Enter and confirm a patient's ID or room at the central station"
to select it.
"Enter and confirm a patient's ID or room at the central station"
"Enter and confirm a patient's ID or room at the central station"
Welch Allyn Acuity Central Monitoring System