Welch Allyn Acuity Directions For Use Manual page 199

Central monitoring systems
Hide thumbs Also See for Acuity:
Table of Contents

Advertisement

Directions for use
Review windows
Table 39. Review window
Features
Frequency of trend data
sampling and storage by Acuity
System
Frequency of patient monitor
data storage and transfer to
Acuity System
Time delay from Virtual Monitor
display to Graphical Trends
Review window display
Display priority of trends
Time span
Cursor bar position
HR/PR, RR/BR, P1/P2, P3/P4,
and CO
labels to the right of the
2
Graphical Trend Display
Presets values
Blue cursor bar shows
24-hour (or up to 96-hour) time
line indicates patient and
equipment state
Printouts
Graphical Trends
Customizable default
Eight hours
Chapter 11 Specifications
Factory default
Sampled and stored once per second
Data is stored at the monitor every two minutes (one
minute for Propaq LT) and then transferred to Acuity
System
Less than one minute, updated continuously
Up to five vital signs (ten in the full-screen review
view) that have been active within last 24 hours, in
following order of priority:
1. HR/PR
2. PVC
3. ST
4. P1
5. P2
6. P3
7. P4
8. CO
2
9. RR/BR
10. IPI
11. SpO
2
12. SpO
- PI
2
13. NIBP
14. Temp
The most recent time monitored appears at the right
edge of the window, or if another review window
was viewed first, the period is centered around the
cursor bar position in the first-viewed window
Reflect labels selected at the monitor during the
current time position of the cursor bar.
P1 is 30 mmHg to 180 mmHg; P2, P3, and P4 are 0
mmHg to 50 mmHg; all others similar to monitor
factory "Trend Graphs" values. For further
information, see the monitor directions for use.
Vital sign numerics and event types. Event markers
are arrhythmia types, Arrhythmia and ST Relearn or
Learning Done and snapshot print.
Time line color matches icon color at all points in
time
Graph, Table, Summary
195