Welch Allyn Acuity Directions For Use Manual page 146

Central monitoring systems
Hide thumbs Also See for Acuity:
Table of Contents

Advertisement

142
Chapter 9 Status icons
Table 8. Icons and messages that represent patient and monitor status
Patient/monitor
status icon
Monitor
Communication Lost
Alert Tone Suspended
Icon display
Icon location on map
priority
Take priority
Locations representing
over low-level
problem monitors.
alerts.
During
Locations representing
suspension
monitors with suspended
period, newly
audible alert tones.
occurring alerts
are indicated
only after
original alert is
corrected.
Acuity System
only visually
indicates newly
occurring
alarms by
changing the
Alert Tone
Suspended icon
from yellow to
red.
Welch Allyn Acuity Central Monitoring System
Messages in
Recommended action
hallway message
panels and at top
of Acuity Central
Station screen
[room #]
If the disconnect is
Loss of Comms
intentional, click Confirm,
and then close the alert
window. Otherwise,
reestablish
communications between
the patient monitor and
the Acuity System.
[room #] Equipment
Attend to problem. Then
or more specific
click the Silence/Reset
message
key at the patient monitor
or Resume at the central
station Virtual Monitor.