Directions for use
Table 35. Setup window
Features
Alarm level
Arrhythmia alarm limits
Presets values
Print on alarm
Options:
Irregular Limit
Single ECG (active with 5-wire cable
only)
Pacer Display
Analyze Pacers
PaceMaker NonCapture
Leads used for arrhythmia analysis
Arrhythmia alarms for paced patients
Arrhythmia Alarms
—
Customizable default
Medium for Pause, off for all
other arrhythmia types except
VFib, asystole and V-Tach
V-Tach, 125; PVC run, 6; PVC/
min, 30; Tachycardia, 180;
Bradycardia, 45; Irregular Limit,
45
Off for all
Off
Off
Chapter 11 Specifications
Factory default
Lethal for VFib, asystole, V-Tach
Presets values
Off
Off
Disabled
Welch Allyn 1500 Patient Monitor,
Micropaq, and Propaq LT monitor 5-wire
cables:
• Arrhythmia channel 0 uses lead II.
• Arrhythmia channel 1 uses lead V.
• Arrhythmia channel 2 uses lead III.
Propaq Encore, Propaq CS:
The leads that the user selects for ECG1
and ECG2 are used for arrhythmia
analysis.
Arrhythmia alarms depend on Acuity
System pacemaker detection vs. monitor
pacemaker detection.
191