Welch Allyn Acuity Directions For Use Manual page 164

Central monitoring systems
Hide thumbs Also See for Acuity:
Table of Contents

Advertisement

160
Chapter 10 Troubleshooting and maintenance
CO
, SpO
, IPI, and NIBP displays
2
2
Table 16. Troubleshooting CO
Problem
When I try to change CO
the patient monitor, a message
says I must use the Acuity System
units.
When I select SpO
rate source, ECG remains the
source.
SpO
or NIBP are not displayed in
2
the Virtual Monitor when I try to
turn on either for display.
A connected monitor equipped
with IPI does not display an IPI
value in the Virtual Monitor.
An NIBP reading seems
inaccurate.
"??:??:??" is shown to indicate the
time of an NIBP measurement.
When NIBP measurements are
taken while a monitor is
temporarily disconnected from an
Acuity System, the NIBP values
are not transferred to the Acuity
System review windows.
, SpO
, IPI, and NIBP displays
2
2
Explanation
units at
CO
units are an Acuity System installation choice set by your system administrator.
2
2
Acuity System requires a single CO
correctly.
as the pulse
When ECG is monitored and C-LOCK is on, ECG is chosen as the heart rate source.
2
You can select up to three waveforms for display. If you select more than three, the
three highest priority waveforms are displayed.
ECG is always displayed when it is monitored, and P1, P2, P3, P4, and CO
priority over other selections. If you want to display SpO
P4, and CO
.
2
In addition, NIBP numerics display for only one hour after an NIBP measurement is
taken.
This occurs under these circumstances:
• The monitor's Patent Mode is set to neonate. Make sure that the monitor is set
to adult or pediatric mode.
• Not all parameters used to calculate IPI are active. Make sure that etCO
Respiration Rate, SpO
WARNING If a noninvasive blood pressure reading is suspect, repeat the
measurement, watching a manometer (if the monitor has one) to verify the
reading. If you are still uncertain about the reading, use another method.
This occurs under either of these circumstances:
• The monitor was connected to the Acuity System during NIBP measurement,
and the monitor was then disconnected from and reconnected to the Acuity
System.
• The monitor was not connected to the Acuity System during NIBP
measurement, and then the monitor was connected to the Acuity System.
Open the NIBP status window at the patient monitor to read the time of the NIBP
value.
If an NIBP measurement is taken within two minutes before connection to an Acuity
System or two minutes following disconnection from an Acuity System, the reading
might not transfer to the Acuity System.
Take NIBP measurements after the monitor is connected to the Acuity System, to
ensure NIBP trending.
Welch Allyn Acuity Central Monitoring System
unit to ensure that the CO
2
or NIBP, turn off P1, P2, P3,
2
,and Pulse Rate are are being monitored.
2
units are displayed
2
take
2
,
2

Advertisement

Table of Contents
loading

This manual is also suitable for:

Mobile acuity lt

Table of Contents