Welch Allyn Acuity Directions For Use Manual page 53

Central monitoring systems
Hide thumbs Also See for Acuity:
Table of Contents

Advertisement

Directions for use
Adjust preferences for a patient's central station printouts
This section presents the Printouts Setup window, in which you can specify details
regarding automatic central station printouts.
To adjust preferences for a patient's printouts
1. In the patient's Virtual Monitor, click Setup, Printouts.
The Printouts Setup window appears.
2. Set preferences in these areas of the window:
Note
These open the Alarms, Printouts, Waveform Window and Arrhythmia Alarms and ST Analysis Setup
windows.
Central Alarm Print: Check a vital-sign box to set automatic printing of an alarm
print if an alarm occurs for that vital-sign.
Print on Alarm Delay: Move the slider to set the time delay (in seconds) between
an alarm and the onset of an alarm printout.
If the duration of an alarm is shorter than this delay period, no alarm print is
generated.
Central Auto Print: Click an interval for automatic snapshot printouts (Auto Prints)
to occur.
Central OxyCRG Alarm Print: Check a vital-sign box to set automatic printing of an
OxyCRG print if an alarm occurs for that vital sign.
OxyCRG printouts combine HR/PR, CO
On some monitors, RR/BR is labeled RRECG/RRCO2. RR and RRECG are
respiration rates derived from ECG, while BR and RRCO2 are respiration rates
derived from CO
.
2
Chapter 4 Adjust monitoring settings and patient information
, RR/BR and SpO
2
49
Prints a Snapshot printout at
the Acuity Central
Station.
vital signs, if active.
2

Advertisement

Table of Contents
loading

This manual is also suitable for:

Mobile acuity lt

Table of Contents